FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE

MDR report key: 7927309 · Received October 2, 2018

Report

Report Number
1920898-2018-00759
Event Type
Malfunction
Date Received
October 2, 2018
Date of Event
September 9, 2018
Report Date
October 15, 2018
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: CUSTOMER RETURNED (1) LOOSE 1CC, 12.7MM SYRINGE. CUSTOMER STATES THAT A PIECE OF BLACK HAIR FOUND INSIDE THE BARREL. THE RETURNED SYRINGE WAS EXAMINED AND EXHIBITED A DARK STRAND OF MATERIAL INSIDE THE BARREL. A SMALL PORTION OF THIS MATERIAL WAS REMOVED FROM THE SAMPLE AND PREPARED FOR FTIR SPECTRAL ANALYSIS. THE ANALYSIS SHOWS THAT THIS MATERIAL IS MOST LIKELY A STRAND OF HAIR MIXED WITH SILICONE. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH# 7241945. ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE FIVE (5) NOTIFICATIONS (B)(4) NOTED THAT DID NOT PERTAIN TO THE COMPLAINT. BASED ON THE SAMPLES / PHOTO(S) RECEIVED THE INVESTIGATION CONCLUDED: CONFIRMED: BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE. PROBABLE ROOT CAUSE IS LIKELY TO BE INEFFECTIVE GOWNING PROCEDURE BY AN ASSOCIATE DURING MANUFACTURING OF THIS PRODUCT. WHEN THIS TYPE OF EVENT OCCURS, HUMAN HAIRS AND DEBRIS MAY FIND IT'S WAY INTO THE PROCESS/PRODUCT WITHOUT DETECTION, RESULTING IN INSTANCES SUCH AS THOSE NOTED WITH THE RETURNED SAMPLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THERE WAS "A PIECE OF BLACK HAIR" FOUND INSIDE THE BARREL OF THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS "A PIECE OF BLACK HAIR" FOUND INSIDE THE BARREL OF THE BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE. THERE WAS NO REPORT OF EXPOSURE, INJURY, OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
768310 BD INSULIN SYRINGE WITH THE BD ULTRA-FINE¿ NEEDLE PISTON SYRINGE FMF BD MEDICAL - DIABETES CARE 7241945

Patients

Seq Age Sex Outcome Treatment
1 Other