FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBES

MDR report key: 7926118 · Received October 2, 2018

Report

Report Number
9617032-2018-02826
Event Type
Malfunction
Date Received
October 2, 2018
Date of Event
September 7, 2018
Report Date
January 28, 2019
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
UDI-DI
50382903630481
PMA / PMN Number
K013971
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION/TESTING AND UPON COMPLETION, THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILL WAS NOT OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. INVESTIGATION CONCLUSION: BASED ON EVALUATION OF THE RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILL WITH THE INCIDENT LOT WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. ROOT CAUSE DESCRIPTION: A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBES UNDER FILLED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8011715; MEDICAL DEVICE EXPIRATION DATE: 2018-09-30; DEVICE MANUFACTURE DATE: 2018-01-11. MEDICAL DEVICE LOT #: 8004784; MEDICAL DEVICE EXPIRATION DATE: 2018-10-31; DEVICE MANUFACTURE DATE: 2018-01-04. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBES UNDER FILLED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
767462 BD VACUTAINER® 9NC 0.109M BUFFERED TRISODIUM CITRATE PLUS BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) SEE H.10 50382903630481

Patients

Seq Age Sex Outcome Treatment
1 Other