PELVIC ARRAY ADAPTOR ASSY
Report
- Report Number
- 3005985723-2018-00567
- Event Type
- Malfunction
- Date Received
- October 2, 2018
- Date of Event
- September 6, 2018
- Report Date
- December 5, 2018
- Manufacturer
- MAKO SURGICAL CORP.
- Product Code
- OLO
- UDI-DI
- 00848486002930
- PMA / PMN Number
- K121064
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
REPORTED EVENT: IT WAS REPORTED THAT ADAPTER TIGHTENING BOLT BROKE OFF IN 3 PIN CLAMP. PRODUCT EVALUATION AND RESULTS: THE PRODUCT WAS UNAVAILABLE FOR INSPECTION AS THE PRODUCT WAS NOT RETURNED. PRODUCT HISTORY REVIEW: REVIEW OF THE PRODUCT HISTORY RECORDS INDICATE 46 DEVICES WERE MANUFACTURED UNDER LOT NO 19130711 AND ACCEPTED INTO FINAL STOCK ON (B)(6) 2011. NO NON-CONFORMANCES WERE IDENTIFIED DURING INSPECTION. COMPLAINT HISTORY REVIEW: A REVIEW OF COMPLAINTS IN CATSWEB AND TRACKWISE RELATED TO P/N 112240, LOT NUMBER 19130711, SHOWS 01 ADDITIONAL COMPLAINTS RELATED TO THE FAILURE IN THIS INVESTIGATION. COMPLAINT (B)(4). CONCLUSIONS: THE FAILURE MODE COULD NOT BE CONFIRMED BECAUSE THE PART WAS NOT AVAILABLE FOR EVALUATION. IF DEVICE AND/OR ADDITIONAL INFORMATION BECOME AVAILABLE, THIS INVESTIGATION WILL BE REOPENED. CORRECTIVE ACTION/PREVENTIVE ACTION: NO ACTION IS REQUIRED AT THIS TIME AS THERE IS NO INDICATION TO SUGGEST A PRODUCT NON-CONFORMITY OR UNANTICIPATED HAZARD.
ADAPTER TIGHTENING BOLT BROKE OFF IN 3 PIN CLAMP. CASE TYPE: THA.
AS PART OF NORMAL COMPLAINT FOLLOW-UP, AN EVALUATION OF THE EVENT HAS BEEN INITIATED BY MAKO SURGICAL. A SUPPLEMENTAL REPORT WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
ADAPTER TIGHTENING BOLT BROKE OFF IN 3 PIN CLAMP. CASE TYPE: THA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 769739 | PELVIC ARRAY ADAPTOR ASSY | STEREOTAXIC DEVICE, ROBOTICS | OLO | MAKO SURGICAL CORP. | 19130711 | 00848486002930 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |