SENSOR ENLITE MMT-7008A
Report
- Report Number
- 2032227-2018-35218
- Event Type
- Malfunction
- Date Received
- September 28, 2018
- Date of Event
- July 14, 2018
- Report Date
- February 5, 2020
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSPECTED ONE OPENED/USED SENSOR AND PERFORMED CONTINUITY RESISTANCE TEST AND SENSOR FAILED TEST. FOUND CANNULA BENT UNABLE TO CONFIRM CUSTOMER RECEIVED SENSOR IN SAID CONDITION DUE TO CUSTOMER RETURNED OPENED/USED.
THE CUSTOMER REPORTED VIA PHONE CALL THE SENSOR HAD INACCURATE READINGS THAT TRIGGERED THRESHOLD SUSPEND ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 122 MG/DL AND THE SENSOR GLUCOSE WAS 60 MG/DL. CUSTOMER OTHER RELEVANT BLOOD GLUCOSE LEVEL WAS 140, 155 AND 122 MG/DL. INSULIN DELIVERY WAS SUSPENDED DUE TO SENSOR VALUES. THE CUSTOMER WAS INFORMED THAT THEIR BLOOD GLUCOSE AND SENSOR GLUCOSE LEVELS WERE NOT IN ACCEPTABLE RANGE. THE SENSOR WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 758486 | SENSOR ENLITE MMT-7008A | ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND | OZO | MEDTRONIC MINIMED | MMT-7008A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 27 YR |