T:SLIM G5 SYSTEM
Report
- Report Number
- 3013756811-2018-34534
- Event Type
- Malfunction
- Date Received
- September 27, 2018
- Date of Event
- September 12, 2018
- Report Date
- September 27, 2018
- Manufacturer
- TANDEM DIABETES CARE
- Product Code
- OYC
- UDI-DI
- 00853052007257532071
- PMA / PMN Number
- P140015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- US
- Reporter Occupation
- OTHER
Narratives
THE X2 USER GUIDE STATE ¿HUMALOG® OR NOVOLOG® . ONLY HUMALOG® AND NOVOLOG® HAVE BEEN TESTED AND FOUND TO BE COMPATIBLE FOR USE IN THE SYSTEM. USE OF INSULIN WITH LESSER OR GREATER CONCENTRATION CAN RESULT IN UNDER DELIVERY OR OVER DELIVERY OF INSULIN. THIS CAN CAUSE VERY HIGH OR A VERY LOW BLOOD GLUCOSE.¿ PER TANDEM¿S USER GUIDE: ¿NEVER FILL YOUR TUBING WHILE YOUR INFUSION SET IS CONNECTED TO YOUR BODY. ALWAYS ENSURE THAT THE INFUSION SET IS DISCONNECTED FROM YOUR BODY BEFORE FILLING THE TUBING. FAILURE TO DISCONNECT YOUR INFUSION SET FROM YOUR BODY BEFORE FILLING THE TUBING CAN RESULT IN OVER DE¬LIVERY OF INSULIN. THIS CAN CAUSE SERIOUS INJURY OR DEATH FROM VERY LOW BLOOD GLUCOSE.¿ NO PRODUCT WAS RETURNED FOR EVALUATION. SHOULD NEW RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT MULTIPLE MINIMUM FILL NOTIFICATIONS OCCURRED AFTER FILLING MULTIPLE CARTRIDGES WITH 120-140 UNITS OF INSULIN. ADDITIONALLY, IT WAS REPORTED THAT THE CUSTOMER FILLED TUBING WHILE CONNECTED AT THE SITE. REPORTEDLY, A NEW CARTRIDGE WAS LOADED SUCCESSFULLY AND INSULIN DELIVERY WAS RESUMED. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL REACHED 221-301 MG/DL.
CUSTOMER LOADED HUMALIN U100 INTO THE CARTRIDGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 754939 | T:SLIM G5 SYSTEM | CONTINUOUS GLUCOSE MONITOR | OYC | TANDEM DIABETES CARE | 1000096 | M022071 | 00853052007257532071 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |