FDA Adverse Event Other Summary report: N

BD PARADIGM LINK BLOOD GLUCOSE MONITOR

MDR report key: 791515 · Received November 15, 2006

Report

Report Number
2243072-2006-00411
Event Type
Other
Date Received
November 15, 2006
Date of Event
October 15, 2006
Report Date
November 15, 2006
Manufacturer
BECTON DICKINSON
Product Code
NBW
PMA / PMN Number
k041478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

VERIFICATION OF THE ACCURACY FOR THIS LOT WAS PERFORMED AGAINST THE CURRENT VERSION OF THE INSPECTION TEST PROCEDURE. LOT PASSED ALL CRITERIA OUTLINED IN THE TEST PROCEDURE. WILL CONTINUE TO MONITOR ON MONTHLY COMPLAINT TREND REPORTS.

Description of Event or Problem · 1

CONSUMER CALLED TO REPORT ERRATIC READINGS WITH THE BGM METER. ANALYSIS RUN ON CONSENSUS ERROR GRID USING MEAN READING (300) AS REFERENCE POINT VS. BD READINGS (230, 103, 569) YIELDED SOME DATA POINTS IN THE C ZONE OF THE GRID, OUTSIDE OF ACCEPTABLE RANGES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BD PARADIGM LINK BLOOD GLUCOSE MONITOR BD TEST STRIPS NBW BECTON DICKINSON NA 6064067

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other