ID-MTS A/B/D MONOCLONAL & REVERSE GROUPING CARD
Report
- Report Number
- 1056600-2006-00346
- Event Type
- Malfunction
- Date Received
- November 27, 2006
- Date of Event
- October 31, 2006
- Report Date
- November 25, 2006
- Manufacturer
- MICRO TYPING SYSTEMS, INC.
- Product Code
- KSZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- INVALID DATA
Narratives
SAMPLE WAS SUBMITTED FOR INVESTIGATION. CUSTOMER COMPLAINT WAS NOT VERIFIED AT MTS. SAMPLE IS A WEAK D EXAMPLE REACTING CONSISTENTLY IN ANTI-D GEL AND POSITIVE IN TUBE AT AHG PHASE ONLY AT MTS. MTS IS UNABLE TO EXPLAIN FALSE NEGTIVE REACTIONS OBSERVED BY THE CUSTOMER IN THE ANTI-D MICROTUBE. ALTHOUGH INCIDENT APPEARS TO BE SAMPLE RELATED, MTS IS UNABLE TO RULE OUT GEL CARD MALFUNCTION AND/OR USER ERROR AS CONTRIBUTING FACTORS. LABELING CAUTIONS THE USER AS FOLLOWS: VERY WEAK EXPRESSIONS OF THE D ANTIGEN MAY NOT BE DETECTED BY THE MTS MONOCLONAL ANTI-D GEL CARD. IN INSTANCES WHERE CONFIRMATION OF D NEGATIVE ANTIGEN STATUS IS REQUIRED, NEGATIVE D REACTIONS OBTAINED WITH THE MTS MONOCLONAL ANTI-D SHOULD BE RETESTED WITH AN ANTI-D REAGENT LICENSED FOR ANTIGLOBULIN PHASE TESTING. LESS THAN OPTIMAL TEST CELL CONCENTRATIONS MAY RESULT IN TEST REACTIONS, WHICH CANNOT BE FULLY OBSERVED. A D+ PATIENT ERRONEOUSLY TYPED AS D NEGATIVE: (1) WOULD RECEIVE D NEGATIVE BLOOD AND (2) MIGHT RECEIVE ANTI-D IMMUNOGLOBULIN DURING OR AFTER PREGNANCY OR AFTER TRANSFUSION OF D POSITIVE BLOOD TRANSFUSION OF D NEGATIVE BLOOD PRODUCTS TO A D POSTIVE PATIENT IS OF NO CLINICAL SIGNIFICANCE. THE UNNECESSARY RECEIPT OF ANTI-D IMMUNOGLOBULIN CARRIES A RISK OF ADVERSE REACTIONS. THE LIKELIHOOD OF A SERIOUS INJURY IS REMOTE IN THIS SITUATION. A FALSE NEGATIVE RESULT WOULD LEAD TO WITHHOLDING OF ANTI-D IMMUNOGLOBULIN IN A NON-RHD IMMUNIZED MOTHER AND SUBSEQUENT RISK (1-10%) OF D IMMUNIZATION. THE SEVERITY LEVEL IS SEVERE/LIFE THREATENING DUE TO THE CUMULATIVE POTENTIAL RISKS OF A FALSE NEG. ALTHOUGH SERIOUS TRANSFUSION REACTIONS DUE TO ANTI-D ARE NOT USUAL, THEY ARE PROBABLY NOT REMOTE.
CUSTOMER REPORTED THAT A PATIENT'S SAMPLE TESTED WITH MTS A/B/D MONOCLONAL AND REVERSE GROUPING CARD LOT# 022806037-32 FAILED TO REACT IN THE ANTI-D MICROTUBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ID-MTS A/B/D MONOCLONAL & REVERSE GROUPING CARD | A/B/D MONOCLONAL & REVERSE GROUPING CARD | KSZ | MICRO TYPING SYSTEMS, INC. | * | 022806037-32 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |