REVACLEAR
Report
- Report Number
- 3006552611-2018-00014
- Event Type
- Injury
- Date Received
- September 25, 2018
- Report Date
- November 12, 2018
- Manufacturer
- BAXTER HEALTHCARE - OPELIKA
- Product Code
- KDI
- UDI-DI
- 07332414101701
- PMA / PMN Number
- K060195
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NURSE
Narratives
THE DEVICE HAS BEEN RECEIVED, THE SAMPLE WAS ANALYZED FOR PVP/NMP CONTENT PER PROTOCOL AND THE RESULTS WERE WITHIN EXPECTATION. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE INITIAL ONSET WAS REPORTED AS ON (B)(6) 2018 AND THEN ON (B)(6) 2018 AND LASTLY THE (B)(6) 2018. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT A PATIENT UNDERGOING TREATMENT WITH A REVACLEAR MAX DIALYZER EXPERIENCED HYPERSENSITIVITY REACTIONS. ON THE DATE OF ONSET OF REACTIONS, NO MEDICAL INTERVENTION WAS REPORTED FOR THIS OCCURRENCE. TWO DAYS LATER AND 15 MINUTES INTO TREATMENT WITH THE SAME DIALYZER, THE PATIENT EXPERIENCED A ¿FIRST ANAPHYLACTIC REACTION¿ AND WAS LATER HOSPITALIZED THAT SAME DAY. THE PATIENT WAS TREATED WITH AND AN EPI-PEN (EPINEPHRINE INJECTION). TWO DAYS LATER, THE PATIENT WHILE UNDERGOING TREATMENT WITH THE SAME DEVICE, EXPERIENCED A "SEVERE ANAPHYLACTIC REACTION". THE PATIENT'S SYMPTOMS AT TIME OF REACTION RANGED FROM ¿HIVES, CLOSING OF THROAT, DIFFICULT BREATHING, WEAVING¿. AN EPI-PEN WAS ADMINISTERED AND THE SYMPTOMS DECREASED. AT THE TIME OF THIS REPORT, THE PATIENT WAS STILL HOSPITALIZED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 749446 | REVACLEAR | DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM | KDI | BAXTER HEALTHCARE - OPELIKA | NA | C418205501 | 07332414101701 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Hospitalization| R | REVACLEAR DIALYZER |