FCS CERAMIC 54 OR 56X28
Report
- Report Number
- 1818910-2018-70382
- Event Type
- Injury
- Date Received
- September 25, 2018
- Date of Event
- January 1, 2018
- Report Date
- September 3, 2018
- Manufacturer
- DEPUY INTERNATIONAL LTD. 801037
- Product Code
- MRA
- PMA / PMN Number
- P040023
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. 'PRODUCT COMPLAINT #: (B)(4). INVESTIGATION SUMMARY: THE DEVICES WERE REVIEWED BY BIOENGINEERING AND A REPORT WAS RECEIVED STATING IT IS UNLIKELY THAT A POTENTIAL PRODUCT ISSUE WAS PRESENT ROOT CAUSE UNDETERMINED. DEPUY SYNTHES CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION MAY BE RE-OPENED AS NECESSARY. DEVICE HISTORY LOT: PRODUCT CODE 159963054, LOT NUMBER 1160577 WAS MANUFACTURED ON DEC 2003. (B)(4) PARTS WERE MANUFACTURED PER SPECIFICATION AND ALL RAW MATERIALS MET SPECIFICATION. THERE WERE NO NCS OR DEVIATIONS ASSOCIATED WITH THIS LOT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
PRODUCT COMPLAINT #: (B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE COMPLAINT DATABASE SEARCH FOUND NO OTHER REPORTED INCIDENT(S) AGAINST THE PROVIDED PRODUCT/LOT COMBINATION(S) SINCE RELEASE FOR DISTRIBUTION. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
REVISION DUE TO SOFT TISSUE INJURY, INSTABILITY, IMPLANT FRACTURE, NOISE, AND DISLOCATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 748506 | FCS CERAMIC 54 OR 56X28 | DURALOC OPTION IMPLANTS : HIP CERAMIC ACETABULAR LINERS | MRA | DEPUY INTERNATIONAL LTD. 801037 | 1160577 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR | Required Intervention |