FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA 1500

MDR report key: 7905768 · Received September 25, 2018

Report

Report Number
2517506-2018-00548
Event Type
Malfunction
Date Received
September 25, 2018
Date of Event
August 30, 2018
Report Date
September 25, 2018
Manufacturer
SIEMENS HEATLHCARE DIAGNOSTICS INC.
Product Code
JJE
PMA / PMN Number
K051087
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC) AND REPORTED THAT A DISCORDANT, FALSELY LOW GLU RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE CCC SPECIALIST REVIEWED THE INSTRUMENT FILES AND DETERMINED THAT QUALITY CONTROL RESULTS RECOVERED WITHIN EXPECTED RANGES ON THE DAY OF THE EVENT. A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER'S SITE. THE CSE RAN SERVICE METHOD TESTS (SMT) ON ALIGNMENTS ON SAMPLE PROBE 1, REAGENT PROBE 1, AND REAGENT PROBE 2, REALIGNED S1 AND R1, REPEATED SMT ON ALIGNMENT FOR S1 AND R1, AND RAN MIX TEST SMT FOR S1, R1 AND R2. THE MIX TEST AND ALIGNMENT SMT TESTS RECOVERED WITHIN SPECIFICATIONS. THEN, THE CSE RAN QUALITY CONTROLS (QCS), RESULTING ACCEPTABLY. THE CAUSE OF THE DISCORDANT, FALSELY LOW GLU RESULT IS UNKNOWN. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A DISCORDANT, FALSELY LOW GLUCOSE (GLU) RESULT WAS OBTAINED ON A PATIENT SAMPLE ON A DIMENSION VISTA 1500 INSTRUMENT. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLES WERE REPEATED TWICE ON THE SAME INSTRUMENT , RESULTING HIGHER. THE REPEAT RESULT OF 82 MG/DL WAS REPORTED AS THE CORRECT RESULT, TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY LOW GLU RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
748345 DIMENSION VISTA 1500 CLINICAL CHEMISTRY ANALYZER JJE SIEMENS HEATLHCARE DIAGNOSTICS INC. DIMENSION VISTA 1500

Patients

Seq Age Sex Outcome Treatment
1