FDA Adverse Event Malfunction Summary report: N

BD SEDITAINER¿ BLOOD COLLECTION TUBES

MDR report key: 7905476 · Received September 25, 2018

Report

Report Number
9617032-2018-02809
Event Type
Malfunction
Date Received
September 25, 2018
Date of Event
September 4, 2018
Report Date
October 19, 2018
Manufacturer
BECTON, DICKINSON AND COMPANY (BD)
Product Code
JKA
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING TO CLOTTING AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BASED ON EVALUATION OF THE CUSTOMER AND RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE PRODUCT WAS FOUND TO BE IN CONFORMANCE AND MEET RELEASE SPECIFICATIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED AFTER USE OF THE BD SEDITAINER BLOOD COLLECTION TUBE THE BLOOD HAS CLOTTED IN THE TUBE. THERE WAS NO REPORT OF EXPOSURE OR MEDICAL INTERVENTION NOTED.

Additional Manufacturer Narrative · 1

MEDICAL DEVICE LOT #: BATCH NUMBER [3080736] WAS NOT FOUND FOR (B)(4). (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED AFTER USE OF THE BD SEDITAINER¿ BLOOD COLLECTION TUBE THE BLOOD HAS CLOTTED IN THE TUBE. THERE WAS NO REPORT OF EXPOSURE OR MEDICAL INTERVENTION NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
746014 BD SEDITAINER¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON AND COMPANY (BD) UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 Other