BD SEDITAINER¿ BLOOD COLLECTION TUBES
Report
- Report Number
- 9617032-2018-02809
- Event Type
- Malfunction
- Date Received
- September 25, 2018
- Date of Event
- September 4, 2018
- Report Date
- October 19, 2018
- Manufacturer
- BECTON, DICKINSON AND COMPANY (BD)
- Product Code
- JKA
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE WITH THE INCIDENT LOT WAS NOT OBSERVED. ADDITIONALLY, RETENTION SAMPLES WERE SELECTED FROM BD INVENTORY FOR EVALUATION AND UPON COMPLETION, NO ISSUES WERE OBSERVED RELATING TO CLOTTING AS ALL SAMPLES MET SPECIFICATIONS. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BASED ON EVALUATION OF THE CUSTOMER AND RETAIN SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL SAMPLES MET THE REQUIRED SPECIFICATIONS. BASED ON THE INVESTIGATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE PRODUCT WAS FOUND TO BE IN CONFORMANCE AND MEET RELEASE SPECIFICATIONS. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. THE BD BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED AFTER USE OF THE BD SEDITAINER BLOOD COLLECTION TUBE THE BLOOD HAS CLOTTED IN THE TUBE. THERE WAS NO REPORT OF EXPOSURE OR MEDICAL INTERVENTION NOTED.
MEDICAL DEVICE LOT #: BATCH NUMBER [3080736] WAS NOT FOUND FOR (B)(4). (B)(6). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED AFTER USE OF THE BD SEDITAINER¿ BLOOD COLLECTION TUBE THE BLOOD HAS CLOTTED IN THE TUBE. THERE WAS NO REPORT OF EXPOSURE OR MEDICAL INTERVENTION NOTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 746014 | BD SEDITAINER¿ BLOOD COLLECTION TUBES | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON AND COMPANY (BD) | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |