FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART XL+ DEFIBRILLATOR/MONITOR
MDR report key: 7897719
·
Received September 21, 2018
Report
- Report Number
- 1218950-2018-07532
- Event Type
- Malfunction
- Date Received
- September 21, 2018
- Report Date
- August 29, 2018
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MKJ
- PMA / PMN Number
- K110825
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED TO PHILIPS THAT THE DEVICE HAD A HIF PCA PROBLEM. THERE WAS NO REPORTED PATIENT OR USER INVOLVEMENT AND NO ADVERSE PATIENT/USER IMPACT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 740279 | HEARTSTART XL+ DEFIBRILLATOR/MONITOR | ALS DEFIBRILLATOR MONITOR | MKJ | PHILIPS MEDICAL SYSTEMS | 861290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |