FDA Adverse Event Malfunction Summary report: N

CELL-DYN 22 CALIBRATOR

MDR report key: 789702 · Received December 1, 2006

Report

Report Number
2919069-2006-00097
Event Type
Malfunction
Date Received
December 1, 2006
Date of Event
October 3, 2006
Report Date
October 3, 2006
Manufacturer
ABBOTT LABORATORIES
Product Code
KRZ
PMA / PMN Number
K902892/A
Removal / Correction Number
2919069-9/26/06-003-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON-MARKET INSTABILITY FOR PLATELETS WITH CELL-DYN 22 CALIBRATOR LOTS 3098 AND 3099 IS BEING EXPERIENCED. THE CALIBRATOR COULD READ HIGHER BY AT LEAST 10% ON THE CELL-DYN RUBY AND AS MUCH AS 22% ON THE CELL-DYN 3200. IF THIS CALIBRATOR IS USED AND THE CALIBRATION FACTOR ON THE INSTRUMENT IS ADJUSTED, A NEGATIVE PLATELET SAMPLE BIAS COULD RESULT. AS A PRECAUTION, LOT 3100 WAS ALSO REMOVED FROM USE. INVESTIGATION INTO THE CAUSE OF THIS INSTABILITY IS IN PROGRESS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 1

THE CUSTOMER REQUESTED REPLACEMENT OF CELL-DYN 22 CALIBRATOR LOT 3100 WHICH WAS DISCARDED PER FA18SEP2006 CUSTOMR LETTER. THERE IS NO REPORT OF INCORRECT PATIENT RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CELL-DYN 22 CALIBRATOR CD22 CALIBRATOR KRZ ABBOTT LABORATORIES NA 3100

Patients

Seq Age Sex Outcome Treatment
1 NO INFO CELL-DYN 3200 ANALYZER LIST 4H59-03,