CELL-DYN 22 CALIBRATOR
Report
- Report Number
- 2919069-2006-00113
- Event Type
- Malfunction
- Date Received
- December 1, 2006
- Report Date
- September 28, 2006
- Manufacturer
- ABBOTT DIAGNOSTICS DIVISON/CELLDYN
- Product Code
- KRZ
- PMA / PMN Number
- k902892/a
- Removal / Correction Number
- 2919069-9/26/06-003-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MS, US
- Reporter Occupation
- OTHER
Narratives
ON-MARKET INSTABILITY FOR PLATELETS WITH CELL-DYN 22 CALIBRATOR LOTS 3098 AND 3099 IS BEING EXPERIENCED. THE CALIBRATOR COULD READ HIGHER BY AT LEAST 10% ON THE CELL-DYN RUBY AND AS MUCH AS 22% ON THE CELL-DYN 3200. IF THIS CALIBRATOR IS USED AND THE CALIBRATION FACTOR ON THE INSTRUMENT IS ADJUSTED, A NEGATIVE PLATELET SAMPLE BIAS COULD RESULT. AS A PRECAUTION, LOT 3100 WAS REMOVED FROM USE. INVESTIGATION INTO THE CAUSE OF THIS INSTABILITY IS IN PROGRESS. THIS IS AN INITIAL REPORT. AN INVESTIGATION IS IN PROCESS. A FINAL REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
THE ACCOUNT STATED THAT THEIR PLATELET CONTROLS ARE TRENDING HIGH ON THE CD3200 SL ANALYZER SINCE THEY CALIBRATED WITH CELL-DYN 22 CALIBRATOR LOT #3099. THERE WAS NO REPORT OF SUSPECT PATIENT RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CELL-DYN 22 CALIBRATOR | CD22 CALIBRATOR | KRZ | ABBOTT DIAGNOSTICS DIVISON/CELLDYN | NA | 3099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | CELL-DYN 3200 SL ANALYZER LIST # 4H60-01 |