FDA Adverse Event Malfunction Summary report: N

BD SYRINGE LUER-LOK¿ TIP

MDR report key: 7894480 · Received September 20, 2018

Report

Report Number
1213809-2018-00617
Event Type
Malfunction
Date Received
September 20, 2018
Date of Event
August 20, 2018
Report Date
October 15, 2018
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903029953
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE LOOSE 10ML ASSEMBLED SYRINGE WAS RECEIVED AND REPORTED TO BE FROM BATCH #8150893 (P/N 302995). THE SAMPLE WAS VISUALLY EVALUATED. THE SYRINGE WAS FOUND TO HAVE A CRACK ALONG THE BARREL WALL OUTSIDE THE PRINT AREA. THE CRACK EXTENDED FROM THE BARREL ROOF TO APPROXIMATELY THE 9ML MARKING. THE CRACK OBSERVED IS REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE DAMAGED BARREL IS ASSOCIATED WITH THE ASSEMBLY PROCESS. NO CORRECTIVE ACTIONS ARE NECESSARY BASED ON THE DEFECTIVE RATE IDENTIFIED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT A BD SYRINGE LUER-LOK¿ TIP HAD A CRACKED SYRINGE. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY OR MEDICAL INTERVENTION.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A BD SYRINGE LUER-LOK¿ TIP HAD A CRACKED SYRINGE. THERE WAS NO REPORT OF EXPOSURE, SERIOUS INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
734866 BD SYRINGE LUER-LOK¿ TIP SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 8150893 30382903029953

Patients

Seq Age Sex Outcome Treatment
1 Other