FDA Adverse Event Injury Summary report: N

PLATE, FIXATION, BONE

MDR report key: 7894200 · Received September 20, 2018

Report

Report Number
8030965-2018-56497
Event Type
Injury
Date Received
September 20, 2018
Report Date
August 28, 2018
Manufacturer
OBERDORF SYNTHES PRODUKTIONS GMBH
Product Code
HRS
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WERE 213 MALES AND 46 FEMALES WITH AN AVERAGE AGE OF 39 YEARS OLD (RANGE, 13-73 YEARS). DATE OF EVENT: UNKNOWN. THIS REPORT IS FOR AN UNKNOWN SYNTHES 3.5MM RECONSTRUCTION PLATE OR 3.5MM SMALL FRAGMENT LOCKING COMPRESSION PLATE/UNKNOWN LOT. PART AND LOT NUMBER ARE UNKNOWN; UDI NUMBER IS UNKNOWN. IMPLANTED DATE: UNKNOWN. EXPLANTED DATE: UNKNOWN. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. (B)(4).

Description of Event or Problem · 1

THIS REPORT IS BEING FILED AFTER THE REVIEW OF THE FOLLOWING JOURNAL ARTICLE: MEEUWIS MA, PULL TER GUNNE AF, VERHOFSTAD MHJ, VAN DER HEIJDEN FHWM (2017). CONSTRUCT FAILURE AFTER OPEN REDUCTION AND PLATE FIXATION OF DISPLACED MIDSHAFT CLAVICULAR FRACTURES. INJURY, INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED. VOLUME 48. ISSUE 3. PAGES 715-719. (THE NETHERLANDS). THE PRIMARY AIM OF THIS RETROSPECTIVE STUDY WAS TO GIVE A DESCRIPTION OF CONSTRUCT FAILURE AFTER PLATE FIXATION OF MIDSHAFT CLAVICULAR FRACTURES. ITS SECONDARY AIM WAS TO IDENTIFY POSSIBLE RISK FACTORS FOR CONSTRUCT FAILURE INCLUDING PATIENT CHARACTERISTICS, FRACTURE TYPE, IMPLANT TYPE AND FRACTURE FIXATION METHOD. BETWEEN JANUARY 1, 2007 AND DECEMBER 31, 2015, A TOTAL OF 259 PATIENTS WITH FRESH MIDSHAFT CLAVICULAR FRACTURE TREATED WITH OPEN REDUCTION AND INTERNAL FIXATION USING A 3.5-MM RECONSTRUCTION PLATE (LOCKING COMPRESSION DESIGN) OR A .3.5-MM PRESHAPED OR NONPRESHAPED SMALL FRAGMENT LOCKING COMPRESSION PLATE, WERE INCLUDED IN THE STUDY. THERE WERE 213 MALES AND 46 FEMALES WITH AN AVERAGE AGE OF 39 YEARS OLD (RANGE, 13-73 YEARS). 50 CLAVICULAR FRACTURES WERE FIXATED USING AN UNKNOWN SYNTHES 3.5MM RECONSTRUCTION PLATE AND 209 FRACTURES WITH AN UNKNOWN SYNTHES 3.5MM SMALL FRAGMENT LOCKING COMPRESSION PLATE. PATIENTS WERE FOLLOWED UP AT LEAST 2 WEEKS, 6 WEEKS (WITH RADIOGRAPHIC CONTROL) AND 3 MONTHS (WITH RADIOGRAPHIC CONTROL) AFTER SURGERY AND CONTINUED UNTIL THE FRACTURE WAS COMPLETELY CONSOLIDATED. COMPLICATIONS WERE REPORTED AS FOLLOWS: (THE ARTICLE DID NOT SPECIFY WHICH PLATES WERE IMPLANTED FOR THE UNKNOWN PATIENTS INVOLVED) ONE HUNDRED AND TWENTY-FOUR PATIENTS HAD REVISION FOR REMOVAL OF PLATES DUE TO IRRITATION. THREE PATIENTS HAD REVISION DUE TO NONUNION. TWO PATIENTS HAD REVISION WITH GENTAMICIN BEADS DUE TO INFECTION. THIS REPORT IS FOR AN UNKNOWN SYNTHES 3.5MM RECONSTRUCTION PLATE OR 3.5MM SMALL. FRAGMENT LOCKING COMPRESSION PLATE. THIS IS REPORT 4 OF 5 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
735911 PLATE, FIXATION, BONE HRS OBERDORF SYNTHES PRODUKTIONS GMBH

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention