FDA Adverse Event Malfunction Summary report: N

COAGULATION FACTOR IX DEFICIENT PLASMA

MDR report key: 7890954 · Received September 19, 2018

Report

Report Number
9610806-2018-00098
Event Type
Malfunction
Date Received
September 19, 2018
Date of Event
August 6, 2018
Report Date
November 9, 2018
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH
Product Code
GJT
UDI-DI
00842768011023
PMA / PMN Number
K924396
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION (20-SEPTEMBER-2018): THE CUSTOMER CLARIFIED THAT THE RESULTS OBTAINED ON NOVEMBER 01, 2017 WERE USED FOR TROUBLESHOOTING PURPOSES. SIEMENS DETERMINED THAT THE INSTRUMENTS AND REAGENTS USED WITHIN THE DIFFERENT TESTS AND QUALITY CONTROLS (QCS) WERE ACCEPTABLE. SIEMENS DETERMINED THAT THE GENETIC DISORDER OF THE PATIENT POTENTIALLY CONTRIBUTED TO THE TESTS TO REACT DIFFERENTLY. CORRECTED INFORMATION (21-SEPTEMBER-2018): THE AFFECTED FACTOR IX (FIX) LOT IS 504129B. THE REAGENT AND INSTRUMENT ARE PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THIS DEVICE IS REQUIRED. MDRS 9610806-2018-00094_S1, 9610806-2018-00095_S1, 9610806-2018-00096_S1, AND 9610806-2018-00097_S1 WERE FILED FOR THE SAME EVENT.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION (24-OCTOBER-2018): ON 24-OCT-2018, SIEMENS WAS NOTIFIED THAT CUSTOMER REPORTED THE EVENT TO THE FDA. THE CUSTOMER ALSO INDICATED THAT THE PATIENT HAS HISTORY OF BLEEDING AFTER HYPOSPADIAS REPAIR AND A TONSILLECTOMY AT AGE (B)(6). MDRS 9610806-2018-00094_S2, 9610806-2018-00095_S2, 9610806-2018-00096_S2, AND 9610806-2018-00097_S2 WERE FILED FOR THE SAME EVENT.

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED A SIEMENS CUSTOMER CARE CENTER (CCC) AND REPORTED THAT A DISCORDANT, FALSELY ELEVATED COAGULATION FACTOR IX DEFICIENT PLASMA (FIX),RESULTS WERE OBTAINED ON 4 SAMPLES FROM THE SAME PATIENT ON THE BCS XP SYSTEM. THERE WAS NO INDICATION OF A SYSTEM MALFUNCTION AND THERE WERE NO REPORTS OF HARDWARE ERRORS WHEN THE DISCORDANT RESULTS WERE OBTAINED ON THE PATIENT SAMPLES. QUALITY CONTROLS WERE WITHIN EXPECTED RANGES. CUSTOMER ALSO RAN NON-SIEMENS ASSAYS ON THE SIEMENS BCS XP SYSTEM, RESULTING IN USER DEFINED METHOD OF TESTING. SIEMENS IS INVESTIGATING THE ISSUE. MDR 9610806-2018-00094, MDR 9610806-2018-00095, AND MDR 9610806-2018-00096 WERE FILED FOR THE DISCORDANT RESULTS OBTAINED ON (B)(6) 2017 AND (B)(6) 2017 FOR THE SAME ISSUE. MDR 9610806-2018-00097 WAS FILED FOR DISCORDANT RESULTS OBTAINED ON (B)(6) 2018 FOR THE SAME ISSUE.

Description of Event or Problem · 1

DISCORDANT, FALSELY ELEVATED COAGULATION FACTOR IX (FACTOR IX) RESULTS WERE OBTAINED ON A PATIENT SAMPLE ON A BCS XP SYSTEM, USING THE COAGULATION FACTOR IX DEFICIENT PLASMA REAGENT. THE DISCORDANT RESULT WAS NOT REPORTED TO THE PHYSICIAN(S). THE SAMPLE WAS REPEATED USING A NON-SIEMENS METHOD AND ON A BCS XP SYSTEM, USING THE SAME COAGULATION FACTOR IX DEFICIENT PLASMA REAGENT LOT AND A NON-SIEMENS REAGENT. THE RESULTS OBTAINED USING THE NON-SIEMENS REAGENT RECOVERED LOWER THAN MOST RESULTS OBTAINED USING THE SIEMENS REAGENT. THE SAMPLE WAS ALSO SENT TO AN ALTERNATE LAB, WHERE IT WAS REPEATED USING A NON-SIEMENS METHOD, RESULTING LOWER. THE LOWER RESULTS MATCHED THE CLINICAL PICTURE FOR THE PATIENT. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED FACTOR IX RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
730872 COAGULATION FACTOR IX DEFICIENT PLASMA COAGULATION FACTOR IX DEFICIENT PLASMA GJT SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH COAGULATION FACTOR IX DEFICIENT PLASMA 504129B 00842768011023

Patients

Seq Age Sex Outcome Treatment
1 16 YR