FDA Adverse Event Malfunction Summary report: N

BD LUER -LOK¿ SYRINGE WITH DETACHABLE NEEDLE

MDR report key: 7890426 · Received September 19, 2018

Report

Report Number
8041187-2018-00361
Event Type
Malfunction
Date Received
September 19, 2018
Date of Event
August 9, 2018
Report Date
September 24, 2018
Manufacturer
BECTON DICKINSON MEDICAL (SINGAPORE)
Product Code
FMI
UDI-DI
00382903057795
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: ONE RETURNED SAMPLE RECEIVED FOR INVESTIGATION. THE RETURNED SAMPLE WAS SUBJECTED TO VISUAL INSPECTION AND HUB LEAK TEST. BASED ON THE VISUAL INSPECTION AND HUB LEAK TEST, THE SAMPLE PASSED BOTH ACCEPTANCE CRITERIA. DURING THE VISUAL INSPECTION AND HUB LEAK TEST, THERE WAS NO CLOGGED NEEDLE OBSERVED. A LOT HISTORY REVIEW WAS CARRIED OUT AND NO RELATED NON CONFORMANCE'S WERE RAISED IN ASSOCIATION WITH THIS PACKAGED LOT CONCLUDING ALL INSPECTIONS WERE PERFORMED PER THE APPLICABLE OPERATIONS AND MET QC SPECIFICATIONS. INVESTIGATION CONCLUSION: BASED ON THE INVESTIGATION FINDING, WE ARE NOT ABLE TO CONFIRM THE DEFECT. AS SUCH, WE ARE UNABLE TO DETERMINE THE ROOT CAUSE.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD LUER -LOK¿ SYRINGE WITH DETACHABLE NEEDLE LEAKED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. PMA/510(K)#: K980987(SYRINGE); K161170 (NEEDLE). A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD LUER -LOK¿ SYRINGE WITH DETACHABLE NEEDLE LEAKED. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
731426 BD LUER -LOK¿ SYRINGE WITH DETACHABLE NEEDLE SYRINGE FMI BECTON DICKINSON MEDICAL (SINGAPORE) 7176198 00382903057795

Patients

Seq Age Sex Outcome Treatment
1 Other