FDA Adverse Event Malfunction Summary report: N

COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE

MDR report key: 788830 · Received November 21, 2006

Report

Report Number
2023826-2006-01781
Event Type
Malfunction
Date Received
November 21, 2006
Report Date
October 23, 2006
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
p990013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

H6: EVALUATION CODES: RESULTS - (OTHER): VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND ONE HAPTIC TORN OFF AND MISSING. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSIONS - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015 COLLAMER THREE PIECE LENS AND THE HAPTIC TORE. THE LENS WAS REMOVED WITH NO PATIENT INJURY. ADDITIONAL INFORMATION HAS BEEN REQUESTED, BUT TO DATE NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THIS IS ONE OF FOUR LENSES USED ON THIS PATIENT - SEE MFR. REPORT #2023826-2006-01782, #2023826-2006-01783, #2023826-2006-01784.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE INTRAOCULAR LENS HQL STAAR SURGICAL CO. CQ2015 NA

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN INJECTOR MODEL MSI-PM - LOT NUMBER UNK.| CARTRIDGE MODEL CQ CARTRIDGE-FP - LOT NUMBER UNK.