COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE
Report
- Report Number
- 2023826-2006-01784
- Event Type
- Malfunction
- Date Received
- November 21, 2006
- Report Date
- October 23, 2006
- Manufacturer
- STAAR SURGICAL CO.
- Product Code
- HQL
- PMA / PMN Number
- p990013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
H6. EVALUATION RESULTS - VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A LARGE PIECE OF THE LENS OPTIC AND ONE HAPTIC TORN OFF AND MISSING. THERE WAS EVIDENCE OF CLEAR SURGICAL RESIDUE. CONCLUSION - (NO CONCLUSION CAN BE DRAWN): BASED ON THE COMPLAINT HISTORY AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. IT SHOULD BE NOTED THAT THE INJECTOR AND CARTRIDGE WERE NOT RETURNED FOR EVALUATION.
IT WAS REPORTED THAT THE SURGEON INSERTED A CQ2015 COLLAMER THREE PIECE LENS AND THE HAPTIC TORE. THE LENS WAS REMOVED WITH NO PT INJURY. ADDITIONAL INFO HAS BEEN REQUESTED, BUT TO DATE NONE HAS BEEN FORTHCOMING. IF ADDITIONAL INFO IS RECEIVED, A SUPPLEMENTAL MEDWATCH WILL BE SENT. THIS IS ONE OF FOUR LENSES USED ON THIS PT - SEE MFR. REPORT # 2023826-2006-01781, # 2023826-2006-01782 AND # 2023826-2006-01783
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLAMER ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER THREE PIECE | INTRAOCULAR LENS | HQL | STAAR SURGICAL CO. | CQ2015 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | CARTRIDGE MODEL CQ CARTRIDGE-FP - LOT NO. UNK| INJECTOR MODEL MSI-PM - LOT NO. UNK |