FDA Adverse Event
Malfunction
Summary report: N
SOPHA NUC CAMERA
MDR report key: 78796
·
Received January 29, 1997
Report
- Report Number
- 78796
- Event Type
- Malfunction
- Date Received
- January 29, 1997
- Date of Event
- November 2, 1996
- Report Date
- November 12, 1996
- Manufacturer
- SOPHA/SMV AMERICA
- Product Code
- IYX
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- IL, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
THE NUCLEAR MEDICINE TECH SET UP THE CAMERA FOR PIN HOLE IMAGES. HER FIRST ATTEMPTED IMAGES WOULD NOT ACQUIRE ANY COUNTS. SHE EXITED THE PROGRAM, CHOSE PROTOCOL AGAIN BUT WHEN SHE CHOSE THE STATIC IMAGE BUTTON, THE CAMERA STARTED ROTATING. THE TECH QUICKLY PRESSED THE RED EMERGENCY STOP BUTTON, BUT THE CAMERA KEPT MOVING. THE TECH RUSHED TO MOVE THE PT AWAY FROM THE CAMERA. THE CAMERA HAD ROTATED ENOUGH TO PUT SOME PRESSURE ON THE PT'S CHEST CAUSING A CHEST CONTUSION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SOPHA NUC CAMERA | NUCLEAR MEDICINE CAMERA | IYX | SOPHA/SMV AMERICA | DS7 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR | Other |