FDA Adverse Event Malfunction Summary report: N

SOPHA NUC CAMERA

MDR report key: 78796 · Received January 29, 1997

Report

Report Number
78796
Event Type
Malfunction
Date Received
January 29, 1997
Date of Event
November 2, 1996
Report Date
November 12, 1996
Manufacturer
SOPHA/SMV AMERICA
Product Code
IYX
Product Problem
Yes
Report Source
User Facility report
Reporter Location
IL, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

THE NUCLEAR MEDICINE TECH SET UP THE CAMERA FOR PIN HOLE IMAGES. HER FIRST ATTEMPTED IMAGES WOULD NOT ACQUIRE ANY COUNTS. SHE EXITED THE PROGRAM, CHOSE PROTOCOL AGAIN BUT WHEN SHE CHOSE THE STATIC IMAGE BUTTON, THE CAMERA STARTED ROTATING. THE TECH QUICKLY PRESSED THE RED EMERGENCY STOP BUTTON, BUT THE CAMERA KEPT MOVING. THE TECH RUSHED TO MOVE THE PT AWAY FROM THE CAMERA. THE CAMERA HAD ROTATED ENOUGH TO PUT SOME PRESSURE ON THE PT'S CHEST CAUSING A CHEST CONTUSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SOPHA NUC CAMERA NUCLEAR MEDICINE CAMERA IYX SOPHA/SMV AMERICA DS7 *

Patients

Seq Age Sex Outcome Treatment
1 11 YR Other