STD BRL LAG SCRW 12.5X115MM
Report
- Report Number
- 3002806535-2018-01082
- Event Type
- Injury
- Date Received
- September 14, 2018
- Date of Event
- July 1, 2018
- Report Date
- December 12, 2018
- Manufacturer
- BIOMET UK LTD.
- Product Code
- HSB
- PMA / PMN Number
- N/A
- Removal / Correction Number
- ZFA-2018-00210
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE PRODUCT RELATING TO THE REPORTED EVENT HAS BEEN RETURNED TO (B)(4) FOR EVALUATION. A REVIEW OF THE MANUFACTURING HISTORY RECORDS FOR THE HIPLOC PLATE FROM THIS EVENT HAS NOT REPORTED ANY ABNORMALITIES OR DEVIATIONS. A REVIEW OF THE COMPLAINT DATABASE FOUND 17 SIMILAR COMPLAINTS REPORTED FOR THIS ITEM CODE. THE RETURNED HIPLOC PLATES HAVE BEEN CONFIRMED AS NON-CONFORMING. THE INTERNAL BARREL ON THE HIPLOC PLATE HAS BEEN CONFIRMED AS UNDERSIZED THROUGH DIMENSIONAL CHECKS. A SUPPLIER CORRECTIVE ACTION REPORT, (B)(4), HAS BEEN RAISED ON SANATMETAL TO INVESTIGATE THEIR MANUFACTURING PROCESSES IN RELATION TO THE NON-CONFORMANCE. A CAPA (B)(4) HAS BEEN RAISED TO INVESTIGATE RECEIPT AND RELEASE OF PRODUCT. PRODUCT HOLD HAS BEEN INITIATED: (B)(4). A HHE HAS BEEN COMPLETED: (B)(4). A FIELD ACTION DETERMINATION HAS BEEN COMPLETED: (B)(4).
HIPLOC PLATE AND SCREW COULD NOT BE ASSEMBLED.
(B)(4). MEDICAL PRODUCT - STD BRL 4-HOLE PLATE 135 DEG , ITEM 2110-135-004, LOT 6192491. REPORT SOURCE, FOREIGN - EVENT OCCURRED IN (B)(6). THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED. MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 3002806535-2018-01083.
THE HIPLOC PLATE HOLE IS TOO SMALL FOR THE SCREW TO GO THROUGH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 720298 | STD BRL LAG SCRW 12.5X115MM | HSB | BIOMET UK LTD. | 2017010792 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |