FDA Adverse Event
Malfunction
Summary report: N
ISOLATOR 10 MICROBIAL BLOOD CULTURE TUBE
MDR report key: 787477
·
Received November 15, 2006
Report
- Report Number
- 787477
- Event Type
- Malfunction
- Date Received
- November 15, 2006
- Date of Event
- October 25, 2006
- Report Date
- November 15, 2006
- Manufacturer
- IVERNESS MEDICAL PROFESSIONAL DIAGNOSTICS
- Product Code
- JSC
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- AR, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
IT TAKES AN EXTRA 1-2 DAYS FOR THE ORGANISMS TO GROW OUT, SINCE THE UNLYSED (WHOLE BLOOD) SPECIMEN JUST SITS ON TOP OF THE CHOCOLATE AGAR MEDIA PLATE. NORMALLY THE LYSED BLOOD CELLS ABSORB INTO THE MEDIA. MICROBIOLOGY SUPERVISOR SUSPECTS THE LYSING AGENT IN THE TUBES IS INACTIVE OR CORRUPT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ISOLATOR 10 MICROBIAL BLOOD CULTURE TUBE | TUBE, BLOOD CULTURE | JSC | IVERNESS MEDICAL PROFESSIONAL DIAGNOSTICS | * | 609215 | |
| 2 | ISOLATOR 10 MICROBIAL BLOOD CULTURE TUBE | TUBE, BLOOD CULTURE | JSC | IVERNESS MEDICAL PROFESSIONAL DIAGNOSTICS | * | 609111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |