FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES

MDR report key: 7873886 · Received September 13, 2018

Report

Report Number
1917413-2018-03449
Event Type
Malfunction
Date Received
September 13, 2018
Date of Event
August 23, 2018
Report Date
October 29, 2018
Manufacturer
BECTON, DICKINSON & CO.
Product Code
JKA
UDI-DI
50382903679626
PMA / PMN Number
K945952
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILLING WITH THE INCIDENT LOT WAS NOT OBSERVED. DRAW TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND ALL SAMPLES DREW WITHIN THE REQUIRED SPECIFICATIONS. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOTS AND NO ISSUES WERE IDENTIFIED. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILLING WITH THE INCIDENT LOTS WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Additional Manufacturer Narrative · 0

BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR UNDERFILLING WITH THE INCIDENT LOT WAS NOT OBSERVED. DRAW TESTING OF THE CUSTOMER SAMPLES WAS PERFORMED AND ALL SAMPLES DREW WITHIN THE REQUIRED SPECIFICATIONS. A REVIEW OF THE MANUFACTURING RECORDS WAS COMPLETED FOR THE INCIDENT LOTS AND NO ISSUES WERE IDENTIFIED. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES OF THIS ISSUE ARE IDENTIFIED. BASED ON EVALUATION OF THE CUSTOMER SAMPLES, THE CUSTOMER¿S INDICATED FAILURE MODE FOR UNDERFILLING WITH THE INCIDENT LOTS WAS NOT OBSERVED. FURTHER INVESTIGATION HAS BEEN INITIATED THROUGH A CAPA. THE INVESTIGATION IS STILL ON-GOING AND IMPROVEMENTS WILL BE MADE AS THE POTENTIAL CAUSES ARE IDENTIFIED. A CAPA HAS BEEN INITIATED TO DOCUMENT FURTHER INVESTIGATION AND ROOT CAUSE ANALYSIS RELATING TO THIS ISSUE. THE INVESTIGATION IS CURRENTLY ON-GOING AND WILL BE UPDATED AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED. BASED ON AN ASSESSMENT OF SEVERITY AND FREQUENCY, IT WAS DETERMINED THAT A CAPA IS REQUIRED AT THIS TIME IN ORDER TO DETERMINE THE ROOT CAUSE ASSOCIATED WITH THIS ISSUE AND IMPLEMENT CORRECTIVE ACTIONS. THE INVESTIGATION IS CURRENTLY ON-GOING AND FURTHER IMPROVEMENTS WILL BE MADE AS THE POTENTIAL ROOT CAUSE(S) ARE IDENTIFIED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER PST GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES, THE TUBES HAD LITTLE OR NO VACUUM. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER PST GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES, THE TUBES HAD LITTLE OR NO VACUUM. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Additional Manufacturer Narrative · 1

MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8061754; MEDICAL DEVICE EXPIRATION DATE: 2019-03-31; DEVICE MANUFACTURE DATE: 2018-03-02; MEDICAL DEVICE LOT #: 8099814; MEDICAL DEVICE EXPIRATION DATE: 2019-04-30; DEVICE MANUFACTURE DATE: 2018-04-09. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING USE OF THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN (LH) BLOOD COLLECTION TUBES, THE TUBES HAD LITTLE OR NO VACUUM. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
713554 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARINN (LH) BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. SEE H.10 50382903679626

Patients

Seq Age Sex Outcome Treatment
1 Other