FDA Adverse Event Malfunction Summary report: N

ILED DUO

MDR report key: 7872146 · Received September 13, 2018

Report

Report Number
9681407-2018-00048
Event Type
Malfunction
Date Received
September 13, 2018
Date of Event
August 14, 2018
Report Date
August 14, 2018
Manufacturer
TRUMPF MEDIZIN SYSTEME GMBH + CO. KG
Product Code
FSY
PMA / PMN Number
K061317
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

IT IS KNOWN THAT IMPROPER INSTALLATION, MAINTENANCE, OR COLLISION CAN LEAD TO THE SPRING ARM COVERS FALLING OFF. THE DEVICE WAS INSPECTED BY A TRUMPF MEDICAL SERVICE TECHNICIAN AND REPAIRED.

Description of Event or Problem · 1

WHILE USER FACILITY STAFF WAS CLEANING A TRUMPF MEDICAL SURGICAL LIGHT SYSTEM, THE PLASTIC STRIP OF A ILED DUO'S SPRING ARM JOINT COVER BECAME SEPARATED AND FELL AFTER A COLLISION. THERE WERE NO INJURIES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
715046 ILED DUO SURGICAL LIGHT FSY TRUMPF MEDIZIN SYSTEME GMBH + CO. KG 4028210

Patients

Seq Age Sex Outcome Treatment
1