FDA Adverse Event Injury Summary report: N

MENTOR SMOOTH ROUND MODERATE PROFILE

MDR report key: 7868963 · Received September 12, 2018

Report

Report Number
1645337-2018-05521
Event Type
Injury
Date Received
September 12, 2018
Date of Event
August 1, 2018
Report Date
August 23, 2018
Manufacturer
MENTOR TEXAS
Product Code
FWM
UDI-DI
00081317001300
PMA / PMN Number
P990075
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

UPON RECEIPT BY MENTOR, THE DEVICE RETURNED WITHOUT FLUID. YELLOW MATERIAL WAS OBSERVED WITHIN THE DEVICE. WHITE MATERIAL WAS OBSERVED ON THE SHELL SURFACE. PRODUCT EVALUATION DISCOVERED A RENT MEASURING APPROXIMATELY 0.3 CM ON THE ANTERIOR ASPECT. MICROSCOPIC EXAMINATION OF THE EDGES OF THE RENT GAVE NO INDICATIONS AS TO CAUSE. NO OTHER ANOMALIES WERE DISCOVERED. THE COMPLAINT WAS CONFIRMED SINCE A RUPTURE WAS FOUND ON THE DEVICE. HOWEVER, AT THIS POINT IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF SUCH DAMAGE OR THE ETIOLOGY OF THE WHITE AND YELLOW MATERIAL FOUND ON THE DEVICE. THE DEVICE HISTORY RECORD (DHR) OF LOT NUMBER 231430 WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THERE IS NO EVIDENCE THAT THE ISSUE IS RELATED WITH MANUFACTURING. BREAST IMPLANTS ARE NOT CONSIDERED LIFETIME DEVICES AND DEFLATION IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR PRODUCT INSERT DATA SHEET. (B)(4).

Additional Manufacturer Narrative · 0

ON 9/17/2018,THE MENTOR FAILURE ANALYSIS LAB HAS RECEIVED THE DEVICE FOR EVALUATION. THE ANALYSIS HAS BEGUN, BUT IS NOT COMPLETE AT THIS TIME. WHEN THE INVESTIGATIONAL ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ON, 9/17/2018, MENTOR BECAME AWARE THAT THE CORRECT DATE OF EVENT WAS (B)(6) 2018. MENTOR ALSO BECAME AWARE THAT THE PATIENT UNDERWENT BILATERAL REMOVAL AND REPLACEMENT WITH (LEFT) MENTOR SMOOTH ROUND MODERATE PROFILE 525CC SALINE CATALOG: 3501685 LOT: 6944105 SN: (B)(4) AND (RIGHT) MENTOR SMOOTH ROUND MODERATE PROFILE 525CC SALINE CATALOG: 3501685 LOT: 7580678 SN: (B)(4). MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 1

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT A (B)(6) -YEAR OLD FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH MENTOR SMOOTH ROUND MODERATE PROFILE 525CC SALINE BREAST PROSTHESIS. DEFLATION OF THE LEFT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT REMOVAL ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
711765 MENTOR SMOOTH ROUND MODERATE PROFILE PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS 231430 00081317001300

Patients

Seq Age Sex Outcome Treatment
1 39 YR Required Intervention