FDA Adverse Event Malfunction Summary report: N

MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL

MDR report key: 7867758 · Received September 12, 2018

Report

Report Number
3002808486-2018-01067
Event Type
Malfunction
Date Received
September 12, 2018
Date of Event
August 24, 2018
Report Date
February 15, 2019
Manufacturer
WILLIAM COOK EUROPE
Product Code
KRD
UDI-DI
00827002202344
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURERS REF# (B)(4). G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI) 400 DANIELS WAY BLOOMINGTON, IN 47404 REGISTRATION NO.: 3005580113. G5) SIMILAR TO DEVICE UNDER 510(K) K150964. (B)(4). INVESTIGATION IS STILL IN PROGRESS.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURERS REF# (B)(4). G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI). 400 DANIELS WAY. BLOOMINGTON, IN 47404. REGISTRATION NO.: 3005580113. SUMMARY OF INVESTIGATIONAL FINDINGS: CONCERNING COIL COULD NOT BE PROPERLY CONNECTED TO TDS AND DETACHED INSIDE A CATHETER. THE INVESTIGATION IS BASED ON EVENT DESCRIPTION. THE COIL WAS NOT RETURNED, ONLY THE TDS ((B)(4)) FROM THE PROCEDURE WAS RETURNED. BASED ON THE LIMITED INFORMATION PROVIDED TO US, IT IS NOT POSSIBLE TO CONCLUDE A MOST LIKELY ROOT CAUSE, BUT INCOMPLETE LOADING OR WIRE ROTATED COUNTERCLOCKWISE DURING INSERTION ARE POSSIBLE CAUSES. NO EVIDENCE WAS FOUND SUGGESTING THAT THE DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER E2016032. WILLIAM COOK EUROPE APS (MANUFACTURER) IS SUBMITTING THIS REPORT ON BEHALF OF COOK MEDICAL INCORPORATED (CMI) (IMPORTER). MANUFACTURER REF# (B)(4). G1) NAME AND ADDRESS FOR IMPORTER SITE: COOK MEDICAL INCORPORATED (CMI). 400 DANIELS WAY. BLOOMINGTON, IN 47404. REGISTRATION NO.: 3005580113. SUMMARY OF INVESTIGATIONAL FINDINGS: RE-INVESTIGATED AS PRODUCT WAS RETURNED. THE INVESTIGATION IS BASED ON THE EVENT DESCRIPTION AND PRODUCT EVALUATION OF THE RETURNED COIL AND TDS (B)(4). A MINOR COMPRESSION OF THE TWO MOST PROXIMAL CRESTS ON THE THREAD WAS OBSERVED. THE COIL WAS ATTACHED TO THE TDS, WITH A BIT OF RESISTANCE, BUT COULD BE ASSEMBLED COMPLETELY. THE THREAD OF THE COIL WAS MEASURED AND FOUND ACCORDING TO SPECIFICATIONS. THE CATHETER USED IN THE PROCEDURE IS UNKNOWN. IT IS NOT POSSIBLE TO CONCLUDE AN EXACT ROOT CAUSE, BUT THE MINOR COMPRESSION IN THE COIL THREAD MAY HAVE CAUSED THE TROUBLED AND INCOMPLETE LOADING AND A SLIGHT, UNINTENDED COUNTERCLOCKWISE WIRE ROTATION DURING INSERTION MAY HAVE CAUSED THE IMPROPERLY LOADED COIL TO PREMATURELY DEPLOY IN THE UNKNOWN CATHETER. NO EVIDENCE WAS FOUND SUGGESTING THAT THE DEVICE WAS NOT MANUFACTURED ACCORDING TO SPECIFICATIONS. COOK MEDICAL WILL CONTINUE TO MONITOR FOR SIMILAR EVENTS. THIS REPORT IS REQUIRED BY THE FDA UNDER 21 CFR PART 803. THIS REPORT IS BASED ON UNCONFIRMED INFORMATION SUBMITTED BY OTHERS. NEITHER THE SUBMISSION OF THIS REPORT NOR ANY STATEMENT MADE IN IT IS INTENDED TO BE AN ADMISSION THAT ANY COOK DEVICE IS DEFECTIVE OR MALFUNCTIONED; THAT A DEATH OR SERIOUS INJURY OCCURRED; OR THAT ANY COOK DEVICE CAUSED OR CONTRIBUTED TO; OR IS LIKELY TO CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY IF A MALFUNCTION OCCURRED.

Description of Event or Problem · 0

DESCRIPTION OF EVENT ACCORDING TO INITIAL REPORTER: THE DEVICE WAS USED FOR EMBOLIZATION TO TREAT PDA BY FEMORAL APPROACH. THE COIL COULD NOT BE CONNECTED WITH THE TDS PROPERLY, SO THE COIL BECAME DETACHED FROM THE TDS INSIDE A CATHETER. SINCE THE PHYSICIAN THOUGHT IT WOULD BE RISKY TO CONTINUE TO USE THE DEVICES, HE REPLACED THEM WITH NEW ONES AND COMPLETED THE PROCEDURE. PATIENT OUTCOME: THERE HAVE BEEN NO ADVERSE EFFECTS TO THE PATIENT REPORTED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION REGARDING THE PATIENT AND/OR EVENT HAS BEEN RECEIVED SINCE THE PREVIOUS MEDWATCH REPORT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
709988 MREYE FLIPPER PDA CLOSURE DETACHABLE EMBOLIZATION COIL KRD DEVICE, EMBOLIZATION, VASCULAR KRD WILLIAM COOK EUROPE E3752135 00827002202344

Patients

Seq Age Sex Outcome Treatment
1