FDA Adverse Event Malfunction Summary report: N

HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY

MDR report key: 7866741 · Received September 11, 2018

Report

Report Number
3007042319-2018-04359
Event Type
Malfunction
Date Received
September 11, 2018
Date of Event
August 28, 2018
Report Date
June 10, 2020
Manufacturer
HEARTWARE, INC.
Product Code
DSQ
PMA / PMN Number
P100047
Removal / Correction Number
Z-1903-2018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: EIGHT (8) BATTERIES, TWO (2) CONTROLLER AC ADAPTERS, AND ONE (1) CONTROLLER DC ADAPTER WERE NOT RETURNED FOR EVALUATION. LOG FILE ANALYSIS REVEALED THAT THE CONTROLLER IN USE DURING THE REPORTED EVENT, CONTAINED A FEATURE THAT RECORDS WHETHER A POWER SOURCE EXPERIENCED A COMMUNICATION ERROR OR A DISCONNECTION WITHIN EACH 15 MINUTE INTERVAL. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS. AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND BATTERIES. AN INTERNAL INVESTIGATION EVALUATED MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: CONTROLLER AC ADAPTER (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 CONTROLLER AC ADAPTER (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 CONTROLLER DC ADAPTER (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H3: YES H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 BATTERY (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 213 H6 FDA CONCLUSION CODE(S): 67 BATTERY (B)(4). H3: YES H6 FDA METHOD CODE(S): 4112, 4114 H6 FDA RESULTS CODE(S): 3213 H6 FDA CONCLUSION CODE(S): 12 MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

THIS SUPPLEMENTAL REPORT (1 OF 95 CORRECTIONS IDENTIFIED) IS BEING SUBMITTED AS A CORRECTION TO THE EVENT DATE TO REFLECT THE DATE OF SERVICING IN THE INITIAL REPORT. THE INITIALLY REPORTED AWARE DATE, REPORTING TIMELINES, REPORTED EVENT AND INVESTIGATION DETAILS REMAIN UNCHANGED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 0

PRODUCT EVENT SUMMARY: ONE (1) CONTROLLER DC ADAPTER WAS RETURNED FOR EVALUATION. FAILURE ANALYSIS OF THE RETURNED CONTROLLER DC ADAPTER REVEALED THAT THE DEVICE PASSED VISUAL INSPECTION AND FUNCTIONAL TESTING. ANALYSIS OF THE DATA LOG FILES REVEALED SEVERAL PREMATURE POWER SWITCHING EVENTS THAT WERE DUE TO MOMENTARY DISCONNECTIONS INVOLVING (B)(4). AS A RESULT, THE REPORTED EVENT WAS CONFIRMED. THE MOST LIKELY ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO MOMENTARY DISCONNECTIONS BETWEEN THE CONTROLLER AND BATTERIES. AN INTERNAL INVESTIGATION EVALUATED MOMENTARY DISCONNECTIONS. ADDITIONAL PRODUCTS: D4: 1440, SERIAL OR LOT#: (B)(6). D10: YES, RETURN DATE: 11-MAR-2019. H3: YES. DEV RTN TO MFR? YES. H6: FDA METHOD CODE(S): 10,4112. H6: FDA RESULTS CODE(S): 213 H6: FDA CONCLUSION CODE(S): 67. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

THE VAD DC ADAPTER WAS REMOVED FROM SERVICE.

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT NOT DIAGNOSIS. THE VENTRICULAR ASSIST SYSTEM IS INDICATED FOR USE AS A BRIDGE TO CARDIAC TRANSPLANTATION AND DESTINATION THERAPY IN PATIENTS WHO ARE AT RISK OF DEATH FROM REFRACTORY END-STAGE LEFT VENTRICULAR HEART FAILURE. THE SYSTEM IS DESIGNED FOR IN-HOSPITAL AND OUT-OF-HOSPITAL SETTINGS. OTHER DEVICES INVOLVED IN THIS EVENT: HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ AC ADAPTER / (B)(4)/ MODEL #: 1430DE UDI #: (B)(4). (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿AC ADAPTER / (B)(4)/ MODEL #: 1430DE (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ DC ADAPTER / (B)(4)/ MODEL #: 1440 UDI #:(B)(4). (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-01-31 UDI #: (B)(4) MFG DATE: 2016-01-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-12-31 UDI #: (B)(4) MFG DATE: 2017-12-12 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2017-01-31 UDI #: (B)(4) MFG DATE: 2016-01-31 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-09-30 UDI #: (B)(4) MFG DATE: 2017-09-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2018-06-30 UDI #: (B)(4) MFG DATE: 2017-06-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY/ (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2015-09-30 UDI #: (B)(4) MFG DATE: 2014-09-30 (B)(4). HEARTWARE VENTRICULAR ASSIST SYSTEM ¿ BATTERY / (B)(4)/ MODEL #: 1650DE / EXPIRATION DATE: 2015-09-30 UDI #: (B)(4) MFG DATE: 2014-09-30 (B)(4). IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

IT WAS REPORTED THAT EIGHT VENTRICULAR ASSIST DEVICE (VAD) BATTERIES, TWO VAD AC ADAPTERS AND ONE VAD DC ADAPTER EXHIBITED POWER SWITCHING. LUBRICATION SERVICING WAS PERFORMED. THE POWER SOURCES REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
707107 HEARTWARE VENTRICULAR ASSIST SYSTEM - BATTERY VENTRICULAR (ASSIST) BYPASS DSQ HEARTWARE, INC. 1650DE

Patients

Seq Age Sex Outcome Treatment
1 MCS UNKNOWN VAD