FDA Adverse Event Malfunction Summary report: N

DIDECO ELECTA PREASSEMBLED SURGICAL WASH SET

MDR report key: 786048 · Received September 6, 2006

Report

Report Number
1718850-2006-00040
Event Type
Malfunction
Date Received
September 6, 2006
Date of Event
July 17, 2006
Report Date
August 8, 2006
Manufacturer
SORIN GROUP ITALIA S.R.L.
Product Code
CAC
PMA / PMN Number
k020647
Removal / Correction Number
PENDING
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

COBE CARDIOVASCULAR INC IS THE DISTRIBUTOR OF A SIMILAR PRODUCT SOLD IN THE US. PRODUCT CATALOG NO. COD.740E/175 IS NOT DISTRIBUTED IN US. HOWEVER, CATALOG NO. (WS175E) IS SOLD IN THE US AND IS SIMILAR TO THAT SOLD IN EUROPE. COD.740E/175 DOES CONTAIN THE COMPONENT THAT IS INVOLVED IN THE US FIELD CORRECTION; THEREFORE, A MEDWATCH REPORT WAS DETERMINED TO BE APPROPRIATE. H3,H6: THE PRODUCT INVOLVED WAS RETURNED AND VISUALLY AND DIMENSIONALLY INSPECTED. THE INSPECTION CONFIRMED THE PRESENCE OF DEBRIS AND FOUND A DEFECTIVE COUPLING ANGLE BETWEEN THE CENTRIFUGE MOUNTING RING AND THE BOTTOM OF THE BOWL. ADD'L EVALUATION WAS PERFORMED ON THE WASH SETS IN THE BRACKET THAT INCLUDED THIS LOT. THAT EVALUATION INCLUDED PERFORMANCE TESTING AND REVEALED THAT IF THE BOWL CANNOT BE INSERTED INTO THE CENTRIFUGE IN ACCORDANCE WITH THE IFU, THERE IS A RISK THAT THE BOWL'S ROTATING SEAL AREA COULD DETERIORATE DURING OPERATION, RELEASING PARTICULATES INTO THE PROCESSED BLOOD. THE LOADING OF WASH SETS FROM THIS BRACKET OF PRODUCT CAN BE MORE DIFFICULT DUE TO A MANUFACTURING PROCESS ERROR THAT RESULTED IN A SLIGHT DEFORMATION IN THE MOUNTING RING ON THE BOWL BOTTOM. THE PROCESS ERROR HAS SUBSEQUENTLY BEEN CORRECTED. COMPLETED A RISK ASSESSMENT THAT LEAD TO A FIELD NOTIFICATION THAT INCLUDED THE PRODUCT SOLD IN THE US. THE FIELD NOTIFICATION IN THE US WAS CONDUCTED BY COBE CARDIOVASCULAR, INC, THE DISTRIBUTOR OF THE US DEVICES. THE LOCAL FDA OFFICE WAS NOTIFIED OF THE FIELD ACTION AND THE REPORTING NUMBER WILL BE PROVIDED WHEN IT IS AVAILABLE.

Description of Event or Problem · 1

METAL DUST IN THE BOWL, THIS BOWL IS ONLY A SAMPLE FROM 3 SYSTEMS WITH THE SAME PROBLEM FROM THE SAME CHARGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIDECO ELECTA PREASSEMBLED SURGICAL WASH SET APPARATUS, AUTOTRANSFUSION CAC SORIN GROUP ITALIA S.R.L. NA 0602030110

Patients

Seq Age Sex Outcome Treatment
1 NA