FDA Adverse Event Malfunction Summary report: N

CATHETER, 28 FR RIGHT ANGLE

MDR report key: 7860053 · Received September 10, 2018

Report

Report Number
3011175548-2018-00988
Event Type
Malfunction
Date Received
September 10, 2018
Date of Event
August 13, 2018
Report Date
September 10, 2018
Manufacturer
ATRIUM MEDICAL
Product Code
GBZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THE THORACIC CATHETER WAS RECEIVED AND INSPECTED. THE BOTTOM SIDE OF THE CATHETER POUCH WAS FOUND TO BE UNSEALED. UPON FURTHER INSPECTION IT WAS DETERMINED THAT THERE WERE NO SIGNS THAT THE POUCH HAD BEEN ATTEMPTED TO BE SEALED. THERE WERE NO WITNESS LINES CREATED BY THE SEALER OR DEFORMATION OF THE FILM CAUSED BY THE HEAT OF THE SEALING PROCESS. THE PVC THORACIC CATHETERS ARE SUPPLIED TO ATRIUM BY A CONTRACT MANUFACTURER (DIVERSATEK HEALTHCARE), WHICH IS RESPONSIBLE FOR THE ASSEMBLY AND PACKAGING PROCESSES. ATRIUM RECEIVES THE PRODUCT AND PERFORMS AN AQL SAMPLING AND INSPECTION OF THE PACKAGED ASSEMBLY. THE INSPECTION INCLUDES VERIFICATION OF APPROPRIATE LABELING AND PACKAGING. REVIEW OF THE RECEIVING INSPECTION RECORDS FOR THIS LOT OF CATHETERS WERE FOUND TO HAVE MET SPECIFICATIONS. BASED ON THE CONFIRMATION OF THE COMPLAINT, ATRIUM MEDICAL HAS ISSUED A SUPPLIER CORRECTIVE ACTION REQUEST TO THE SUPPLIER OF THE CATHETERS. THORACIC CATHETERS ARE STERILE, SINGLE USE, DISPOSABLE DEVICES AND ARE INTENDED TO FACILITATE THE EVACUATION OF AIR AND/OR FLUID FROM THE CHEST CAVITY OR MEDIASTINUM. A PACKAGE THAT IS UNSEALED WOULD BE EXPECTED TO BE NOTICED BEFORE IT WAS ENTERED INTO THE STERILE FIELD, AS IT WAS IN THIS CASE. IT WOULD REPRESENT A DELAY IN TREATMENT FOR THE PATIENT. THE INSTRUCTIONS FOR USE (IFU) STATES DO NOT USE IF THE DEVICE OR PACKAGING IS DAMAGED.

Additional Manufacturer Narrative · 1

WE ARE IN THE PROCESS OF PERFORMING THE INVESTIGATION AND WILL SUBMIT THE FOLLOW-UP REPORT ONCE THE EVALUATION IS COMPLETED.

Description of Event or Problem · 1

UPON RECEIPT, IT WAS DISCOVERED THAT A THORACIC CHEST TUBE WAS UNSEALED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
703850 CATHETER, 28 FR RIGHT ANGLE CATHETER, CHOLANGIOGRAPHY GBZ ATRIUM MEDICAL 8128 ME221275

Patients

Seq Age Sex Outcome Treatment
1