INRATIO
Report
- Report Number
- 2954730-2006-00463
- Event Type
- Malfunction
- Date Received
- August 24, 2006
- Date of Event
- August 18, 2006
- Report Date
- August 18, 2006
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE 08/18/06 INRATIO 3.9, LAB 2.5, MEAN 3.0, CONFIDENCE LIMITS 1.9-4.6; INRATIO 3.4, LAB 2.5, MEAN 2.95, CONFIDENCE LIMITS 1.8-4.2; INRATIO 3.9, LAB 2.5, MEAN 3.2, CONFIDENCE LIMITS 1.9-4.6. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. THE RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE FURTHER TESTING IS NOT REQUIRED AT THIS TIME. INRATIO PRECISION DATA PROVIDED BY END-USER LOT 060090: 08/18/06 FIRST TEST INR = 3.9, SECOND TEST INR = 3.4, THIRD TEST INR = 3.9. MEAN = 3.73; SD = 0.288; %CV = 7.7%. THE %CV IS LESS THAN OR EQUAL TO 20%. PER INTERNAL PROCEDURE, THE PRECISION PASSES THE CRITERIA FOR PRECISION. THE TEST IS CONSIDERED PRECISE AND NO FURTHER TESTING IS REQUIRED AT THIS TIME.
CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB. RESULTS AS FOLLOWS: DATE 08/18/06, INRATIO 3.9, LAB 2.5; INRATIO 3.4, LAB 2.5; INRATIO 3.9, LAB 2.5. CALLER ALLEGED IMPRECISION WITH INRATION. RESULTS AS FOLLOWS: 08/18/06 FIRST TEST INR = 3.9, SECOND TEST INR = 3.4, THIRD TEST INR = 3.9.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | * | 060090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |