METRIX AIR
Report
- Report Number
- 1000113657-2018-00931
- Event Type
- Malfunction
- Date Received
- September 7, 2018
- Date of Event
- August 15, 2018
- Report Date
- September 7, 2018
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- UDI-DI
- 00021292007447
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- OTHER
Narratives
(B)(4). PRODUCT NOT YET RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-20-USER'S TEST STRIP HAD POOR STORAGE(KITCHEN). TEST STRIP UDI# ((B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER ON (B)(4) 2018 IN A FOLLOW-UP CALL TO ENSURE THAT THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN; CUSTOMER STATED THAT THE REPLACEMENT PRODUCT IS WORKING TO THEIR SATISFACTION AND THEY HAVE NOT HAD ANY MEDICAL INTERVENTION OR SYMPTOMS SINCE THE LAST CALL.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TESTS RESULTS OF 216, 240, 199, 231 AND 215 MG/DL. THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 120-150MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL ON (B)(6) 2018, A BACK TO BACK BLOOD TEST WAS NOT PERFORMED BY THE CUSTOMER. THE PRODUCT IS NOT STORED ACCORDING TO SPECIFICATION AND IS STORED IN THE KITCHEN. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 11/30/2018 AND OPEN VIAL DATE IS (B)(6) 2018. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY (CUSTOMER COULD NOT READ DATE/TIME): (B)(6). CUSTOMERS TIME AND DATE CANNOT BE CONFIRMED BECAUSE SHE CANNOT READ THEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 698176 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MU2496 | 00021292007447 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR |