FDA Adverse Event Injury Summary report: N

ASFORA ANTERIOR CERVICAL PLATE SYSTEM

MDR report key: 7853298 · Received September 6, 2018

Report

Report Number
2135141-2018-00003
Event Type
Injury
Date Received
September 6, 2018
Date of Event
August 10, 2018
Report Date
September 6, 2018
Manufacturer
MEDICAL DESIGNS, LLC
Product Code
KWQ
PMA / PMN Number
K143688
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BASED ON THE INFORMATION PROVIDED, REVIEW OF THE SURGICAL TECHNIQUE MANUAL, IFU, CERTIFICATES, AND FMEA, THERE IS NO INDICATION OF DEVICE FAILURE. THE C6 FRACTURE CAUSED BY THE PATIENTS FALL REQUIRED A NEW PLATE SIZE TO BE USED. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS PATIENT COMPLIANCE.

Description of Event or Problem · 1

MEDICAL DESIGNS, LLC (MDLLC) WAS CONTACTED ON 08/17/2018 AND INFORMED OF A REVISION SURGERY THAT HAD TAKEN PLACE ON (B)(6) 2018. THE ORIGINAL SURGERY HAD TAKEN PLACE ON (B)(6) 2018. ON (B)(6) 2018, THE SURGEON IMPLANTED A 2 LEVEL 41 MM PLATE (11-1106-FD2-41) AND TWO (2) 4.0 MM X 16 MM VARIABLE ANGLE LOCKING SCREWS (11-1106-FD7-16) AND FOUR (4) 4.4 MM X 16 MM RESCUE SCREWS (11-1106-FD9-16). FOLLOWING THE INITIAL SURGERY, IT WAS REPORTED THAT THE PATIENT WAS NONCOMPLIANT AND NOT FOLLOWING POST-OP INSTRUCTIONS AND CAME BACK TO THE OFFICE WITH RECURRENT NECK PAIN FOLLOWING A FALL. ACCORDING TO X-RAYS THE PATIENT SUFFERED A C6 COMPRESSION FRACTURE DUE TO THE ACCIDENTAL FALL. THEREFORE, THE PATIENT UNDERWENT A REVISION SURGERY ON (B)(6) 2018 TO ADDRESS THE NEW FRACTURE BY REPLACING THE EXISTING 41 MM PLATE WITH A LONGER 44 MM PLATE. THE OLD PLATE AND SCREWS WERE REMOVED WITHOUT ANY DIFFICULTY AND THE SURGEON IMPLANTED A 2 LEVEL, 44 MM PLATE (11-1106-FD2-44) USING FOUR (4) 4.4 MM X 16 MM RESCUE SCREWS (11-1109-FD9-16) AT C4 AND C6. PATIENT IS SAID TO BE DOING WELL FOLLOWING SURGERY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
690939 ASFORA ANTERIOR CERVICAL PLATE SYSTEM ASFORA ANTERIOR CERVICAL PLATE SYSTEM KWQ MEDICAL DESIGNS, LLC

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention