DIDECO COMPACT AUTOTRANSFUSION KIT
Report
- Report Number
- 1718850-2006-00043
- Event Type
- Malfunction
- Date Received
- September 6, 2006
- Date of Event
- June 28, 2006
- Report Date
- July 3, 2006
- Manufacturer
- SORIN GROUP ITALIA S.R.L.
- Product Code
- CAC
- PMA / PMN Number
- K982650
- Removal / Correction Number
- PENDING
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
THE PRODUCT INVOLVED WAS RETURNED AND VISUALLY AND DIMENSIONALLY INSPECTED. THE INSPECTION FOUND A DEFECTIVE COUPLING ANGLE BETWEEN THE CENTRIFUGE MOUNTING RING AND THE BOTTOM OF THE BOWL. ADDITIONAL EVALUATION WAS PERFORMED ON THE WASH SETS IN THE BRACKET THAT INCLUDED THIS LOT. THAT EVALUATION INCLUDED PERFORMANCE TESTING AND REVEALED THAT IF THE BOWL CANNOT BE INSERTED INTO THE CENTRIFUGE IN ACCORDANCE WITH THE IFU, THERE IS A RISK THAT THE BOWL'S ROTATING SEAL AREA COULD DETERIORATE DURING OPERATION, RELEASING PARTICULATES INTO THE PROCESSED BLOOD. THE LOADING OF WASH SETS FROM THIS BRACKET OF PRODUCT CAN BE MORE DIFFICULT DUE TO A MANUFACTURING PROCESS ERROR THAT RESULTED IN A SLIGHT DEFORMATION IN THE MOUNTING RING ON THE BOWL BOTTOM. THE PROCESS ERROR HAS SUBSEQUENTLY BEEN CORRECTED BY SORIN GROUP. SORIN GROUP COMPLETED A RISK ASSESSMENT THAT LEAD TO A FIELD NOTIFICATION THAT INCLUDED THE PRODUCT SOLD IN THE US. THE FIELD NOTIFICATION IN THE US WAS CONDUCTED BY COBE CARDIOVASCULAR, INC. THE DISTRIBUTOR OF THE US DEVICES. THE LOCAL FDA OFFICE WAS NOTIFIED OF THE FIELD ACTION AND THE REPORTING NUMBER WILL BE PROVIDED WHEN IT IS AVAILABLE.
THEY COULDN'T SET UP THE BOWL, DUE TO THE DISTORTION OF THE BOTTOM RING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DIDECO COMPACT AUTOTRANSFUSION KIT | APPARATUS, AUTOTRANSFUSION | CAC | SORIN GROUP ITALIA S.R.L. | NA | 0601230037 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |