I-STAT CG8+ CARTRIDGE
Report
- Report Number
- 2245578-2018-00228
- Event Type
- Malfunction
- Date Received
- September 6, 2018
- Date of Event
- August 22, 2018
- Report Date
- October 16, 2018
- Manufacturer
- ABBOTT POINT OF CARE
- Product Code
- CHL
- UDI-DI
- 10054749000163
- PMA / PMN Number
- K940918
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
APOC INCIDENT #: (B)(4). THE INVESTIGATION WAS COMPLETED ON 10/10/2018. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. WHILE RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AC (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE), A PREVIOUSLY IDENTIFIED UNRELATED ISSUE FOR THE LOT IS BEING INVESTIGATED UNDER QUALITY RECORD (QR) 2206.
(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
ON (B)(6) 2018, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULT ON A (B)(6) YEAR OLD FEMALE PATIENT WITH CHRONIC LYMPHOCYTIC LEUKEMIA. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE. METHOD: DATE: COLLECTED: TESTED: RESULT: SAMPLE: COBAS: (B)(6) 2018, 12:25, 13:03, 8.20 MMOL/L, PLASMA COLLECTED IN LITHIUM HEPARIN. I-STAT: (B)(6) 2018 , 12:25, 13:12, 7.4 MMOL/L, WB COLLECTED IN LITHIUM HEPARIN. PICCOLO XPRESS: (B)(6) 2018, 12:25, 13:16, >8.5 MMOL/L, WB COLLECTED IN LITHIUM HEPARIN. COBAS: (B)(6) 2018, 12:25, 13:23, 4.10 MMOL/L, SERUM. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 692471 | I-STAT CG8+ CARTRIDGE | CG8+ CARTRIDGE | CHL | ABBOTT POINT OF CARE | NA | W18087 | 10054749000163 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR |