FDA Adverse Event Malfunction Summary report: N

I-STAT CG8+ CARTRIDGE

MDR report key: 7850122 · Received September 6, 2018

Report

Report Number
2245578-2018-00228
Event Type
Malfunction
Date Received
September 6, 2018
Date of Event
August 22, 2018
Report Date
October 16, 2018
Manufacturer
ABBOTT POINT OF CARE
Product Code
CHL
UDI-DI
10054749000163
PMA / PMN Number
K940918
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT #: (B)(4). THE INVESTIGATION WAS COMPLETED ON 10/10/2018. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. WHILE RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AC (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE), A PREVIOUSLY IDENTIFIED UNRELATED ISSUE FOR THE LOT IS BEING INVESTIGATED UNDER QUALITY RECORD (QR) 2206.

Additional Manufacturer Narrative · 1

(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON (B)(6) 2018, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT CG8+ CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT POTASSIUM RESULT ON A (B)(6) YEAR OLD FEMALE PATIENT WITH CHRONIC LYMPHOCYTIC LEUKEMIA. THERE WAS NO ADDITIONAL PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. RETURN PRODUCT IS AVAILABLE. METHOD: DATE: COLLECTED: TESTED: RESULT: SAMPLE: COBAS: (B)(6) 2018, 12:25, 13:03, 8.20 MMOL/L, PLASMA COLLECTED IN LITHIUM HEPARIN. I-STAT: (B)(6) 2018 , 12:25, 13:12, 7.4 MMOL/L, WB COLLECTED IN LITHIUM HEPARIN. PICCOLO XPRESS: (B)(6) 2018, 12:25, 13:16, >8.5 MMOL/L, WB COLLECTED IN LITHIUM HEPARIN. COBAS: (B)(6) 2018, 12:25, 13:23, 4.10 MMOL/L, SERUM. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
692471 I-STAT CG8+ CARTRIDGE CG8+ CARTRIDGE CHL ABBOTT POINT OF CARE NA W18087 10054749000163

Patients

Seq Age Sex Outcome Treatment
1 79 YR