FDA Adverse Event Malfunction Summary report: N

SERVO-U

MDR report key: 7846618 · Received September 5, 2018

Report

Report Number
8010042-2018-00462
Event Type
Malfunction
Date Received
September 5, 2018
Date of Event
August 6, 2018
Report Date
October 29, 2018
Manufacturer
MAQUET CRITICAL CARE AB
Product Code
CBK
PMA / PMN Number
K151814
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Additional Manufacturer Narrative · 0

NO SERVICE HAS BEEN REQUESTED BY THE HOSPITAL AND THE VENTILATOR HAS BEEN PLACED BACK IN CLINICAL USE. NO PARTS HAVE REPORTEDLY BEEN REPLACED. NO VENTILATOR LOGS HAVE BEEN PROVIDED. RECOMMENDATIONS HOW TO HANDLE EXPIRATORY PRESSURE SPIKES HAVE BEEN PROVIDED TO THE USER. THE RECOMMENDATIONS INCLUDE ACTIONS SUCH AS REPLACEMENT OF BACTERIA FILTER TYPES AND VENTILATOR MODE CHANGES UPON RECURRENCE. NO INVESTIGATION HAS BEEN POSSIBLE DUE TO REASONS ABOVE, THEREFORE THE ROOT CAUSE OF THE REPORTED EVENT HAS NOT BEEN DETERMINED BUT IT WAS MOST LIKELY DUE TO NOT OPTIMAL PARAMETER SETTINGS FOR THE ACTUAL PATIENT CONDITION OR AN INCREASED EXPIRATORY RESISTANCE, WHERE ONE POSSIBLE CAUSE IS AN OCCLUDED EXPIRATORY BACTERIA FILTER. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE THE VENTILATOR WAS CONNECTED TO A PATIENT, EXPIRATORY PRESSURE SPIKES WITH A RUBBING SOUND WAS NOTED. THERE WAS NO PATIENT HARM. (B)(4).

Description of Event or Problem · 0

IMPORTER REF. #:(B)(4). MANUFACTURER REF. #: (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
688033 SERVO-U VENTILATOR, CONTINUOUS, FACILITY USE CBK MAQUET CRITICAL CARE AB

Patients

Seq Age Sex Outcome Treatment
1