FDA Adverse Event Malfunction Summary report: N

HIGH DEFINITION LCD MONITOR

MDR report key: 7845191 · Received September 5, 2018

Report

Report Number
8010047-2018-01712
Event Type
Malfunction
Date Received
September 5, 2018
Date of Event
August 16, 2018
Report Date
March 25, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FET
PMA / PMN Number
K102379
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 0

THIS IS A SUPPLEMENTAL REPORT FOR MFR REPORT #8010047-2018-01712. OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) COULD NOT INVESTIGATE THE SUBJECT OEV262H, BECAUSE THE SUBJECT OEV262H IS NOT RETURNED TO OMSC. SINCE OMSC COULD NOT INVESTIGATE THE SUBJECT DEVICE, THE EXACT CAUSE HAS BEEN UNKNOWN; HOWEVER, THE FOLLOWING ARE SUPPOSED TO BE THE CAUSE. THE APPEARED IMAGE ON SUBJECT DEVICE BLACKED OUT BY THE HIGH FREQUENCY NOISE OF THE ELECTROSURGICAL UNIT, BECAUSE THE SUBJECT DEVICE WAS PLACED NEXT TO THE ELECTROSURGICAL UNIT. THE INSTRUCTION MANUAL OF THE OEV262H STATES THE CORRESPONDING METHOD IN CASE OF AN ABNORMALITY.

Additional Manufacturer Narrative · 1

THE SUBJECT OEV262H WAS NOT RETURNED TO OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) FOR EVALUATION. OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) CHECKED THE DEVICE HISTORY RECORD OF THE SUBJECT DEVICE; THERE WAS NO IRREGULARITY FOUND. THERE WERE NO FURTHER DETAILS PROVIDED. IF SIGNIFICANT ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

DURING AN UNSPECIFIED ELECTROSURGERY PROCEDURE USING TWO OR MORE UNSPECIFIED MONITORS, THE IMAGE ON THE MONITOR OEV262H, WHICH WAS PLACED IN FRONT OF THE USER, BLACKED OUT. THE USER MOVED THE PATIENT TO ANOTHER ENDOSCOPY ROOM AND COMPLETED THE PROCEDURE. THE USER FACILITY REPORTED THAT THIS PHENOMENON OCCURRED IN BOTH CASES OF THE ELECTROSURGICAL UNIT: ESG-100 AND AN UNSPECIFIED ELECTROSURGICAL UNIT MADE BY ERBE (B)(4), AND ALSO OCCURRED EVEN WHEN THEY CHANGED THE POWER OUTLET OF THE SUBJECT OEV262H AND THE POWER OUTLET OF THE ELECTROSURGICAL UNIT FOR SEPARATE ONES. FURTHER, THE USER FACILITY REPORTED THAT THEY SET UP THE ISOLATION TRANSFORMER: MB631 FOR THE SUBJECT OEV262H, AND ALSO CONDUCTED POTENTIAL EQUALIZATION BETWEEN THE SUBJECT OEV262H AND THE ELECTROSURGICAL UNIT. HOWEVER, THE PHENOMENON DID NOT IMPROVE. THE USER FACILITY PRESUMED THAT THIS PHENOMENON OCCURRED BECAUSE THE SUBJECT OEV262H WAS PLACED NEXT TO THE ELECTROSURGICAL UNIT. THERE WAS NO PATIENT INJURY REPORTED OTHER THAN BEING MOVED TO ANOTHER ENDOSCOPY ROOM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
687288 HIGH DEFINITION LCD MONITOR HIGH DEFINITION LCD MONITOR FET OLYMPUS MEDICAL SYSTEMS CORP. OEV262H

Patients

Seq Age Sex Outcome Treatment
1