VIDISTAR
Report
- Report Number
- 3011617252-2018-00001
- Event Type
- Malfunction
- Date Received
- September 4, 2018
- Date of Event
- August 6, 2018
- Report Date
- August 6, 2018
- Manufacturer
- HITACHI HEALTHCARE AMERICAS INFORMATICS
- Product Code
- LLZ
- PMA / PMN Number
- K083910
- Removal / Correction Number
- 3011617252-08152018-001
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
OUR FIRM INVESTIGATED THE CUSTOMER COMPLAINT OF INTERMIXED IMAGES AND WERE ABLE TO DETERMINE THAT THE PIECE OF SOFTWARE RESPONSIBLE FOR TRANSFERRING THE IMAGES FROM ONE SYSTEM TO ANOTHER (SSHFS VERSION 2.4) WAS INTERMIXING IMAGES UNDER SPECIFIC CIRCUMSTANCES. IMPLEMENTING CHECKSUM LOGIC TO THE APPLICATION'S LOGS REVEALED WHEN IN THE TRANSFER PROCESS IMAGES WERE INTERMIXED. THIS INDICATED THAT THE CULPRIT WAS A BUG IN THE TRANSFER SOFTWARE (SSHFS VERSION 2.4) AND NOT THE MODALITY, VIDISTAR PACS, NOR THE EHR SYSTEM. THE TRANSFER SOFTWARE (SSHFS) IS A THIRD-PARTY OPEN SOURCE SOFTWARE THAT WAS ORIGINALLY CHOSEN AND VALIDATED FOR THE INTERFACE BY VIDISTAR AND THE EHR SYSTEM. FURTHER INVESTIGATION SHOWED INSTABILITY OF THE MOUNT CREATED BY THE TRANSFER SOFTWARE (SSHFS VERSION 2.4) AT THE LEVELS OF TRAFFIC CREATED BY THE INTERFACE THUS OFFERING AN INCREASED LIKELIHOOD OF THE SPECIFIC CIRCUMSTANCES NECESSARY FOR THE BUG TO PRESENT.
THE VIDISTAR PACS AND DICOM VIEWER SOFTWARE SYSTEM IS USED FOR THE TRANSFER, RECORDING, STORAGE, DISPLAY, AND POST-PROCESSING OF DIGITAL MEDICAL IMAGES, INCLUDING DIGITAL IMAGING AND COMMUNICATIONS IN MEDICINE (DICOM) DATA, MEASUREMENT QUANTIFICATION AND DATA MANIPULATION. VIDISTAR PACS IS SUPPORTING AN INTERFACE THAT PROVIDES THE IMAGES AND MEASUREMENTS TO A REMOTE EHR SYSTEM. SSHFS IS A SECURE FILE SYSTEM CLIENT USED TO MOUNT AND INTERACT WITH DIRECTORIES AND FILES LOCATED ON A REMOTE SERVER. DURING A STUDY TRANSFER TO THE REMOTE EHR SYSTEM, THE PATIENT IMAGES INCLUDED AN IMAGE FROM ANOTHER PATIENT'S STUDY. THE ADDITIONAL IMAGE WAS LABELED CORRECTLY, BUT WAS MISFILED. THE PROBLEM WAS CAUGHT BY THE TECHNICIAN BEFORE THE STUDY WAS INTERPRETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 684793 | VIDISTAR | PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM | LLZ | HITACHI HEALTHCARE AMERICAS INFORMATICS | VIDISTAR | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 45 YR | Other |