FDA Adverse Event Malfunction Summary report: N

VIDISTAR

MDR report key: 7842854 · Received September 4, 2018

Report

Report Number
3011617252-2018-00001
Event Type
Malfunction
Date Received
September 4, 2018
Date of Event
August 6, 2018
Report Date
August 6, 2018
Manufacturer
HITACHI HEALTHCARE AMERICAS INFORMATICS
Product Code
LLZ
PMA / PMN Number
K083910
Removal / Correction Number
3011617252-08152018-001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

OUR FIRM INVESTIGATED THE CUSTOMER COMPLAINT OF INTERMIXED IMAGES AND WERE ABLE TO DETERMINE THAT THE PIECE OF SOFTWARE RESPONSIBLE FOR TRANSFERRING THE IMAGES FROM ONE SYSTEM TO ANOTHER (SSHFS VERSION 2.4) WAS INTERMIXING IMAGES UNDER SPECIFIC CIRCUMSTANCES. IMPLEMENTING CHECKSUM LOGIC TO THE APPLICATION'S LOGS REVEALED WHEN IN THE TRANSFER PROCESS IMAGES WERE INTERMIXED. THIS INDICATED THAT THE CULPRIT WAS A BUG IN THE TRANSFER SOFTWARE (SSHFS VERSION 2.4) AND NOT THE MODALITY, VIDISTAR PACS, NOR THE EHR SYSTEM. THE TRANSFER SOFTWARE (SSHFS) IS A THIRD-PARTY OPEN SOURCE SOFTWARE THAT WAS ORIGINALLY CHOSEN AND VALIDATED FOR THE INTERFACE BY VIDISTAR AND THE EHR SYSTEM. FURTHER INVESTIGATION SHOWED INSTABILITY OF THE MOUNT CREATED BY THE TRANSFER SOFTWARE (SSHFS VERSION 2.4) AT THE LEVELS OF TRAFFIC CREATED BY THE INTERFACE THUS OFFERING AN INCREASED LIKELIHOOD OF THE SPECIFIC CIRCUMSTANCES NECESSARY FOR THE BUG TO PRESENT.

Description of Event or Problem · 1

THE VIDISTAR PACS AND DICOM VIEWER SOFTWARE SYSTEM IS USED FOR THE TRANSFER, RECORDING, STORAGE, DISPLAY, AND POST-PROCESSING OF DIGITAL MEDICAL IMAGES, INCLUDING DIGITAL IMAGING AND COMMUNICATIONS IN MEDICINE (DICOM) DATA, MEASUREMENT QUANTIFICATION AND DATA MANIPULATION. VIDISTAR PACS IS SUPPORTING AN INTERFACE THAT PROVIDES THE IMAGES AND MEASUREMENTS TO A REMOTE EHR SYSTEM. SSHFS IS A SECURE FILE SYSTEM CLIENT USED TO MOUNT AND INTERACT WITH DIRECTORIES AND FILES LOCATED ON A REMOTE SERVER. DURING A STUDY TRANSFER TO THE REMOTE EHR SYSTEM, THE PATIENT IMAGES INCLUDED AN IMAGE FROM ANOTHER PATIENT'S STUDY. THE ADDITIONAL IMAGE WAS LABELED CORRECTLY, BUT WAS MISFILED. THE PROBLEM WAS CAUGHT BY THE TECHNICIAN BEFORE THE STUDY WAS INTERPRETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
684793 VIDISTAR PICTURE ARCHIVING AND COMMUNICATIONS SYSTEM LLZ HITACHI HEALTHCARE AMERICAS INFORMATICS VIDISTAR N/A

Patients

Seq Age Sex Outcome Treatment
1 45 YR Other