BD SYRINGE LUER-LOK¿ TIP
Report
- Report Number
- 1213809-2018-00584
- Event Type
- Malfunction
- Date Received
- August 30, 2018
- Date of Event
- August 10, 2018
- Report Date
- October 4, 2018
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 30382903029953
- PMA / PMN Number
- K980987
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: SINCE NO SAMPLES DISPLAYING THE CONDITION REPORTED ARE AVAILABLE FOR EXAMINATION, WE WERE UNABLE TO FULLY INVESTIGATE THIS INCIDENT, THEREFORE A ROOT CAUSE COULD NOT BE DETERMINED. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBERS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT BD SYRINGE LUER-LOK¿ TIP CONTAINED FOREIGN MATTER DESCRIBED AS "WHITE PARTICLES" IN BLOOD AFTER DRAWING. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. MULTIPLE LOT NUMBERS: THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8150883 MEDICAL DEVICE EXPIRATION DATE: 2023-05-31 DEVICE MANUFACTURE DATE: 2018-05-30. MEDICAL DEVICE LOT #: 8002662 MEDICAL DEVICE EXPIRATION DATE: 2022-12-31 DEVICE MANUFACTURE DATE: 2018-01-02. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT BD SYRINGE LUER-LOK¿ TIP CONTAINED FOREIGN MATTER DESCRIBED AS "WHITE PARTICLES" IN BLOOD AFTER DRAWING. NO SERIOUS INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 672618 | BD SYRINGE LUER-LOK¿ TIP | SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | SEE H.10 | 30382903029953 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |