SERVO-U
Report
- Report Number
- 8010042-2018-00458
- Event Type
- Malfunction
- Date Received
- August 30, 2018
- Date of Event
- August 28, 2018
- Report Date
- October 18, 2018
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K151814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).
THE USER FACILITY PERFORMS SERVICE OF THEIR VENTILATORS BY THEMSELVES AND DID NOT REQUEST ANY SERVICE. IT HAS NOT BEEN COMMUNICATED IF ANY PARTS WERE REPLACED. THE INVESTIGATION OF THE PROVIDED VENTILATOR LOGS CONFIRM THE REPORTED EVENT OF ALARMS FOR LOW O2 CONCENTRATION. THE LOGS DO NOT CONTAIN ANY TECHNICAL ERROR CODES THAT INDICATE ANY VENTILATOR MALFUNCTION AT THE TIME OF THE EVENT. THE VENTILATOR PASSED PRE-USE CHECK SUCCESSFULLY PRIOR TO THE START OF VENTILATION. SINCE NO PARTS WERE RETURNED FOR INVESTIGATION, THE ROOT CAUSE OF THE REPORTED ALARMS HAS NOT BEEN DETERMINED. GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).
IT WAS REPORTED THAT WHEN CONNECTED TO A PATIENT, THE VENTILATOR GENERATED ALARMS FOR LOW O2 CONCENTRATION. THERE WAS NO PATIENT HARM. (B)(4).
IMPORTER REF. #: (B)(4). MANUFACTURER REF. #: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 672585 | SERVO-U | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |