FDA Adverse Event Injury Summary report: N

PINN MAR +4 10D 28IDX46OD

MDR report key: 7832157 · Received August 30, 2018

Report

Report Number
1818910-2018-68097
Event Type
Injury
Date Received
August 30, 2018
Date of Event
August 4, 2018
Report Date
August 4, 2018
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
LPH
UDI-DI
10603295013440
PMA / PMN Number
K033273
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. CONCOMITANT MED PRODUCTS: PINN MAR +4 10D 28IDX46OD; DELTA CER HEAD 12/14 28MM +1.5; UNKNOWN HIP FEMORAL STEM; UNKNOWN HIP ACETABULAR CUP.

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # : (B)(4). INVESTIGATION SUMMARY: EXAMINATION OF THE ACETABULAR LINER DOES FIND EVIDENCE TO SUPPORT THE REPORTED DISASSOCIATION EVENT. PRODUCT ERROR / CONTRIBUTION TO THE EVENT IS NOT IDENTIFIED. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

THE PATIENT HAD PTQ 4 YEARS AGO AND HER INSERT WAS DISPLACED FROM THE ACETABULUM. I FORWARD PHOTOS OF THE PRIMARY EXAMINATION, THE CASE OF THE EVOLUTION OF THE CASE AND ALSO THE BOX WITH THE IMPLANTS FOR ANALYSIS. THE PATIENT DID NOT NEED TO BE HOSPITALIZED FOR THIS REASON, NOR DID SHE HAVE ANY COMPLICATIONS DURING THE SURGERY. PRODUCT INFORMATION: REF. (B)(4), LOT 132163 - ACETABULAR LINER 28MM, REF. (B)(4), LOT 3512554 - CERAMIC FEMORAL HEAD 28 MM 12/14. THE REVIEW WAS PERFORMED ON (B)(6) 2018.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
675254 PINN MAR +4 10D 28IDX46OD PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS LPH DEPUY ORTHOPAEDICS, INC. 1818910 132163 10603295013440

Patients

Seq Age Sex Outcome Treatment
1 49 YR Required Intervention