FDA Adverse Event Malfunction Summary report: N

SYSMEX POCH-100I

MDR report key: 7830713 · Received August 29, 2018

Report

Report Number
1000515253-2018-00013
Event Type
Malfunction
Date Received
August 29, 2018
Date of Event
July 30, 2018
Report Date
February 25, 2019
Manufacturer
SYSMEX CORPORATION
Product Code
GKL
UDI-DI
04987562401994
PMA / PMN Number
K032677
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE PCB IS MADE OF FLAME RESISTANT MATERIALS, BUT EXCESSIVE HEAT FROM THE MALFUNCTION POSES A RISK OF INHALATION OF SMOKE AND/OR HARMFUL VAPORS FROM SCORCHED MATERIALS. THE PCB IS NOT ACCESSIBLE TO THE USER, AND IS GROUNDED, REDUCING THE POTENTIAL FOR HARM DUE TO SHOCK. INVESTIGATION FOR ROOT CAUSE IS ONGOING BY SYSMEX CORPORATION (B)(4).

Additional Manufacturer Narrative · 1

THE SYSMEX POCH-100I INSTRUCTIONS FOR USE (IFU), CHAPTER 7 - SAFETY INFORMATION & WASTE DISPOSAL, SECTION 7.1 SAFETY INFORMATIONS, WARNS THE USER: "IF THE INSTRUMENT EMITS UNUSUAL ODORS OR SMOKE, OR IF THE INSTRUMENT LEAKS: SWITCH OFF IMMEDIATELY AND UNPLUG THE POWER CABLE. CONTACT YOUR SYSMEX SERVICE REPRESENTATIVE." THE USER IMMEDIATELY DISCONNECTED THE POWER. SYSMEX CORPORATION (B)(4) (S-CORP) PERFORMED THE INVESTIGATION. S-CORP IDENTIFIED THE DC/DC CONVERTER MOUNTED ON THE LIQUID CRYSTAL DISPLAY (LCD) CONTROL BOARD (PCB NO. (B)(4)) FAILED, AND REQUESTED AN INVESTIGATION FROM THE SUPPLIER OF THE PCB. DUE TO THE CHARRING OF PCB NO. (B)(4), INVESTIGATION WAS IMPOSSIBLE. ROOT CAUSE WAS UNDETERMINED. S-CORP REPLACED THE ANALYZER. PCB NO. (B)(4) IS NOT ACCESSIBLE TO THE USER, AND IS GROUNDED, REDUCING THE POTENTIAL FOR HARM DUE TO SHOCK. THE PCB IS MADE OF FLAME RESISTANT MATERIALS AND COVERED WITH A METAL SHIELD.

Description of Event or Problem · 1

A USER IN (B)(6) DETECTED A BURNING ODOR EMANATING FROM THE ANALYZER AFTER TURNING THE POWER ON. THE USER IMMEDIATELY DISCONNECTED THE POWER. A SERVICE ENGINEER WAS DISPATCHED AND FOUND A BURNT PRINTED CIRCUIT BOARD (PCB). NO USERS SOUGHT MEDICAL ATTENTION DUE TO THE BURNING ODOR.