FDA Adverse Event Injury Summary report: N

CM TITAMAX EX IMPLANT 4.0X9 MM

MDR report key: 7829490 · Received August 29, 2018

Report

Report Number
3008261720-2018-03700
Event Type
Injury
Date Received
August 29, 2018
Date of Event
March 15, 2018
Report Date
August 29, 2018
Manufacturer
NEODENT - JJGC S.A.
Product Code
DZE
UDI-DI
07898237568843
PMA / PMN Number
K101945
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ES
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2015015. INSTRADENT USA, INC IS SUBMITTING THE REPORT ON BEHALF OF NEODENT - JJGC. CONSIDERING THE SURGICAL METHODOLOGY, IT WAS SEEN THAT THE DENTIST HAS SELECTED AN IMPLANT DESIGN WHICH IS NOT RECOMMENDED FOR THE PATIENT'S BONE QUALITY. MOREOVER, IT WAS SEEN THAT THE DENTIST HAS USED MORE DRILLS THAN RECOMMENDED TO PERFORM THE IMPLANT INSTALLATION.

Additional Manufacturer Narrative · 0

EXEMPTION NUMBER: E2015015. INSTRADENT USA, INC IS SUBMITTING THE REPORT ON BEHALF OF NEODENT - JJGC. CONSIDERING THE SURGICAL METHODOLOGY, IT WAS SEEN THAT THE DENTIST HAS SELECTED AN IMPLANT DESIGN WHICH IS NOT RECOMMENDED FOR THE PATIENT'S BONE QUALITY. MOREOVER, IT WAS SEEN THAT THE DENTIST HAS USED MORE DRILLS THAN RECOMMENDED TO PERFORM THE IMPLANT INSTALLATION.

Description of Event or Problem · 0

(B)(4) ¿ THE DENTIST REPORTED THAT 1 MONTH AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 20#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 40 NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II).

Description of Event or Problem · 0

(B)(4). ¿ THE DENTIST REPORTED THAT 1 MONTH AFTER DENTAL IMPLANT WAS INSTALLED IN INTRAORAL REGION 20#, ITS NON-OSSEOINTEGRATION WAS OBSERVED. IN ADDITION, 40NCM OF PRIMARY STABILITY WAS ACHIEVED AND THE PATIENT PRESENTED INSUFFICIENT BONE QUALITY/QUANTITY (BONE TYPE II).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
670873 CM TITAMAX EX IMPLANT 4.0X9 MM ENDOSSEOUS DENTAL IMPLANT DZE NEODENT - JJGC S.A. 800148931 07898237568843

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention