NITINOL TC ELECTRODE 10CM
Report
- Report Number
- 3006630150-2018-02335
- Event Type
- Malfunction
- Date Received
- August 29, 2018
- Date of Event
- August 15, 2018
- Report Date
- August 29, 2018
- Manufacturer
- COSMAN MEDICAL
- Product Code
- GXI
- UDI-DI
- 00813250013054
- PMA / PMN Number
- K050084
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
QUAVE CLINIC 701 GOLF VIEW DRIVE MEDFORD OR 96080 UNITED STATES ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL: TCN-10 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 10CM QUANTITY:1 MODEL: TCN-10 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 10CM 3M CABLE QUANTITY:1 MODEL: TCN-15 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 15CM QUANTITY:1.
DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES TCN-10 (LOT 013017) AND TCN-10 (LOT 110916) SHOWED THAT THE EPOXY WAS DISCOLORED AND A CHIP OUT. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED. THIS FAILURE MECHANISM IS DUE TO EXTERNAL MECHANICAL FORCE AND THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES. EVALUATION OF TCN-15 (LOT 052715) REVEALED A CHIP OUT. THIS FAILURE MECHANISM IS DUE TO EXTERNAL MECHANICAL FORCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 668308 | NITINOL TC ELECTRODE 10CM | PROBE, RADIOFREQUENCY LESION | GXI | COSMAN MEDICAL | TCN-10 | 013017 | 00813250013054 |