FDA Adverse Event Malfunction Summary report: N

NITINOL TC ELECTRODE 10CM

MDR report key: 7826838 · Received August 29, 2018

Report

Report Number
3006630150-2018-02335
Event Type
Malfunction
Date Received
August 29, 2018
Date of Event
August 15, 2018
Report Date
August 29, 2018
Manufacturer
COSMAN MEDICAL
Product Code
GXI
UDI-DI
00813250013054
PMA / PMN Number
K050084
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

QUAVE CLINIC 701 GOLF VIEW DRIVE MEDFORD OR 96080 UNITED STATES ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL: TCN-10 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 10CM QUANTITY:1 MODEL: TCN-10 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 10CM 3M CABLE QUANTITY:1 MODEL: TCN-15 SERIAL/LOT: (B)(4) DESCRIPTION: NITINOL TC ELECTRODE 15CM QUANTITY:1.

Description of Event or Problem · 1

DEVICE EVALUATION PERFORMED ON THE RETURNED ELECTRODES TCN-10 (LOT 013017) AND TCN-10 (LOT 110916) SHOWED THAT THE EPOXY WAS DISCOLORED AND A CHIP OUT. THE COLOR OF NEWLY CURED EPOXY IS AMBER, LIGHT BROWN. IF THE EPOXY IS SUBJECTED TO TOO MANY AUTOCLAVE CYCLES IT BECOMES BRITTLE AND DISCOLORED. THIS FAILURE MECHANISM IS DUE TO EXTERNAL MECHANICAL FORCE AND THERMAL STRESS ASSOCIATED WITH MULTIPLE AUTOCLAVE CYCLES. EVALUATION OF TCN-15 (LOT 052715) REVEALED A CHIP OUT. THIS FAILURE MECHANISM IS DUE TO EXTERNAL MECHANICAL FORCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
668308 NITINOL TC ELECTRODE 10CM PROBE, RADIOFREQUENCY LESION GXI COSMAN MEDICAL TCN-10 013017 00813250013054

Patients

Seq Age Sex Outcome Treatment
1