SERVO-U
Doing adverse event research by hand?
The Research Assistant searches MAUDE for you, groups reports by failure mode, and cites every record.
Free to try · every answer cites its records
Report
- Report Number
- 8010042-2018-00448
- Event Type
- Malfunction
- Date Received
- August 28, 2018
- Date of Event
- August 2, 2018
- Report Date
- October 31, 2018
- Manufacturer
- MAQUET CRITICAL CARE AB
- Product Code
- CBK
- PMA / PMN Number
- K151814
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERSON: (B)(4).
GETINGE USA SALES, LLC (IMPORTER) IS SUBMITTING THIS REPORT ON BEHALF OF MAQUET CRITICAL CARE AB (MANUFACTURER). REF. EXEMPTION #: E2018003. GETINGE USA SALES, LLC 45 BARBOUR POND DRIVE WAYNE, NJ 07470. CONTACT PERON: (B)(4). THE VENTILATOR WAS INVESTIGATED ON SITE BY OUR FIELD SERVICE ENGINEER (FSE) AND THE O2 GAS MODULE AND THE AIR GAS MODULE WERE REPLACED AND RETURNED FOR INVESTIGATION. THE O2 GAS MODULE AND THE AIR GAS MODULE REGULATES THE INSPIRATORY OXYGEN AND AIR GAS FLOW TO THE PATIENT. THE REPORTED FLUCTUATING O2 CONCENTRATION WAS NOT REPRODUCED DURING TEST OF THE RETURNED O2 GAS MODULE AND THE AIR GAS MODULE IN A VENTILATOR. NO ALARMS WERE GENERATED DURING VENTILATION, NOR ANY DEVIATION DURING TESTS IN THE PRODUCTION TEST SYSTEM. NO DEVICE LOG FILES HAVE BEEN RECEIVED. THE REPORTED PROBLEM COULD NOT BE REPRODUCED, THEREFORE THE CAUSE COULD NOT BE DETERMINED IN THIS INVESTIGATION.
IT WAS REPORTED THAT WHILE CONNECTED TO A PATIENT, THE VENTILATOR ALARMED FOR FLUCTUATING O2 CONCENTRATION. THERE WAS NO PATIENT HARM. (B)(4).
IMPORTER REF. #: (B)(4). MANUFACTURER REF. #:(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 666428 | SERVO-U | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | MAQUET CRITICAL CARE AB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |