FDA Adverse Event Malfunction Summary report: N

CENTURION VISION SYSTEM

MDR report key: 7822425 · Received August 28, 2018

Report

Report Number
2028159-2018-01798
Event Type
Malfunction
Date Received
August 28, 2018
Date of Event
July 31, 2018
Report Date
December 7, 2018
Manufacturer
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K121555
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE COMPANY SERVICE REPRESENTATIVE EXAMINED THE SYSTEM AND WAS ABLE TO REPLICATE THE REPORTED EVENT. THE NONCONFORMING MULTIFUNCTIONAL IN-OUTPUT (MFIO) PRINTED CIRCUIT BOARD (PCB) WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THE NONCONFORMING MULTIFUNCTIONAL IN-OUTPUT (MFIO) PRINTED CIRCUIT BOARD (PCB) WAS RECEIVED AND A VISUAL ASSESSMENT OF THE RETURNED SAMPLE FOUND NO NONCONFORMITIES. WHEN CHECKING THE LMZ COMPONENTS FOR SHORT CIRCUITS, A NONCONFORMING COMPONENT (U11) WAS DISCOVERED. THE SYSTEM WAS MANUFACTURED ON JANUARY 16, 2014. BASED ON QA ASSESSMENT, THE PRODUCT MET SPECIFICATIONS AT THE TIME OF RELEASE. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO THAT THE NONCONFORMING COMPONENTS IN THE MFIO PCB, WHICH ARE UNABLE TO SURVIVE A CONVECTION REFLOW PROCESS DESIGNED FOR THE SOLDER PASTE. AN INTERNAL INVESTIGATION HAS BEEN OPENED TO ADDRESS THIS ISSUE. THE MANUFACTURER INTERNAL REFERENCE NUMBER IS:(B)(4).

Additional Manufacturer Narrative · 1

INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL REPORTABLE INFORMATION BECOMES AVAILABLE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT BEFORE SURGERY, THE SYSTEM SHUT DOWN ON ITS OWN. THE SYSTEM WAS SWITCHED OUT TO PROCEED WITH THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
666112 CENTURION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER NA NA

Patients

Seq Age Sex Outcome Treatment
1 Other