FDA Adverse Event
Death
Summary report: N
CARINA
MDR report key: 7820967
·
Received August 27, 2018
Report
- Report Number
- 3009481053-2018-00036
- Event Type
- Death
- Date Received
- August 27, 2018
- Date of Event
- June 20, 2018
- Report Date
- August 27, 2018
- Manufacturer
- HANDICARE AB
- Product Code
- FSA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
Additional Manufacturer Narrative · 1
NO DEVICE PROBLEM FOUND, THE USER MISS HANDLED THE SLING (NOT MANUFCATURED BY HANDICARE) DURING THE INCIDENT.
Description of Event or Problem · 1
PATIENT FELL OFF FROM THE 4 POINTS HIGH BACK SLING SHEET DURING LIFT BY CARINA 350MM WHILE TRANSFERRING FROM BED (AT HIGHEST POSITION) TO WHEEL CHAIR. PATIENT DECEASED IN HITTING HER HEAD TO THE FLOOR. ONE OF THE SLING STRAP (LEFT SIDE LEG) WAS NOT HOOKED ON THE SLING BAR. ACCORDING TO POLICE INVESTIGATION, THE ACCIDENT OCCURRED UNDER THE MISS HANDLING BY CAREGIVER. CAREGIVER ADMITTED. NO DAMAGES TO THE PATIENT LIFT OR SLING SHEET WERE FOUND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 659974 | CARINA | PATIENT LIFTS | FSA | HANDICARE AB | 60600012 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Death |