AMPLICOR CT/NQ TEST FOR C TRACHOMATIS
Report
- Report Number
- 2243471-2006-00013
- Event Type
- Injury
- Date Received
- November 15, 2006
- Report Date
- October 13, 2006
- Manufacturer
- ROCHE MOLECULAR SYSTEMS, INC.
- Product Code
- MKZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE CUSTOMER PROVIDED ROCHE MOLECULAR SYSTEMS WITH SAMPLES FROM THE SPECIMENS IN QUESTION. ROCHE MOLECULAR SYSTEMS, INC. INVESTIGATED THE CUSTOMER'S ALLEGATIONS AND CONFIRMED THAT THE SPECIMENS TESTED NEGATIVE FOR CHLAMYDIA TRACHOMATIS. FURTHER INVESTIGATION DETERMINED THAT THE SPECIMENS CONTAINED A MUTATED STRAIN OF C. TRACHOMATIS. THIS MUTATION RESULTED IN THE DELETION OF APPROX 350 BASE PAIRS FROM THE REGION WITHIN THE CRYPTIC PLASMID DNA COVERED BY THE TEST'S PRIMERS/AND OR PROBE. RMS IS CONTINUING OUR INVESTIGATION TO BETTER UNDERSTAND THE PREVALENCE AND GEOGRAPHIC DISTRIBUTION OF THE MUTATED STRAIN. THE PACKAGE INSERT WILL ALSO BE MODIFIED TO REFLECT THE RARE, BUT POSSIBLE OCCURRENCE OF MUTATIONS THAT COULD IMPACT THE TEST'S PERFORMANCE.
CUSTOMER REPORTED A SERIES OF SPECIMENS THAT WERE NEGATIVE FOR CHLAMYDIA TRACHOMATIC WITH THE ROCHE MOLECULAR SYSTEMS AMPLICOR CT/NG TEST, BUT POSITIVE WITH AN ALTERNATE METHOD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMPLICOR CT/NQ TEST FOR C TRACHOMATIS | IN VITRO DIAGNOSTIC NUCLEIC ACID TEST | MKZ | ROCHE MOLECULAR SYSTEMS, INC. | * | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |