FDA Adverse Event
Malfunction
Summary report: N
FREESTYLE FLASH
MDR report key: 781899
·
Received November 9, 2006
Report
- Report Number
- 2954323-2006-01265
- Event Type
- Malfunction
- Date Received
- November 9, 2006
- Date of Event
- August 3, 2006
- Report Date
- November 9, 2006
- Manufacturer
- ABBOTT DIABETES CARE INC.
- Product Code
- LFR
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE METER HAS BEEN CONFIRMED TO EXHIBIT THE MEMORY OVERWRITE MALFUNCTION. NO MANUFACTURING PARAMETERS FOUND OUT OF SPEC AND FRESH PANASONIC BATTERY VOLTAGE WAS MEASURED AT 3.096V. DID NOT OBSERVE BATTERY ICON OR BOOKLET ICON WHILE STRIP WAS INSERTED. UNIT OF MEASURE WAS SELECTABLE AND WAS REC'D A MMOL/L. 0300, 0806, AND 0015 ERRORS WERE OBSERVED IN THE ERROR LOG INDICATING BATTERY DROOP. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FMA 05/16/06 LETTER.
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT THE UNIT OF MEASUREMENT SETTING OF THEIR FREESTYLE FLASH METER CHANGED FROM MMOL/L TO MG/DL. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FREESTYLE FLASH | BLOOD GLUCOSE MONITORING SYSTEM | LFR | ABBOTT DIABETES CARE INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN |