FDA Adverse Event Malfunction Summary report: N

FREESTYLE FLASH

MDR report key: 781899 · Received November 9, 2006

Report

Report Number
2954323-2006-01265
Event Type
Malfunction
Date Received
November 9, 2006
Date of Event
August 3, 2006
Report Date
November 9, 2006
Manufacturer
ABBOTT DIABETES CARE INC.
Product Code
LFR
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE METER HAS BEEN CONFIRMED TO EXHIBIT THE MEMORY OVERWRITE MALFUNCTION. NO MANUFACTURING PARAMETERS FOUND OUT OF SPEC AND FRESH PANASONIC BATTERY VOLTAGE WAS MEASURED AT 3.096V. DID NOT OBSERVE BATTERY ICON OR BOOKLET ICON WHILE STRIP WAS INSERTED. UNIT OF MEASURE WAS SELECTABLE AND WAS REC'D A MMOL/L. 0300, 0806, AND 0015 ERRORS WERE OBSERVED IN THE ERROR LOG INDICATING BATTERY DROOP. CUSTOMERS AND RETAILERS HAVE BEEN INFORMED THROUGH THE FMA 05/16/06 LETTER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE UNIT OF MEASUREMENT SETTING OF THEIR FREESTYLE FLASH METER CHANGED FROM MMOL/L TO MG/DL. THERE WAS NO REPORT OF DEATH, SERIOUS INJURY OR MISTREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FREESTYLE FLASH BLOOD GLUCOSE MONITORING SYSTEM LFR ABBOTT DIABETES CARE INC. * *

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN