FDA Adverse Event Malfunction Summary report: N

TABLE TOP STERILIZER

MDR report key: 7818699 · Received August 24, 2018

Report

Report Number
MW5079403
Event Type
Malfunction
Date Received
August 24, 2018
Date of Event
August 8, 2018
Report Date
August 14, 2018
Manufacturer
TUTTNAUER USA CO., LTD
Product Code
FLE
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

DURING REGULAR CYCLE, THE TABLE TOP STERILIZERS FAILED SAFE. EVALUATION OF THEM SHOWED THAT WATER PUMP IS DEFECTIVE. THE PUMP IS FURNISHED AS PART OF STERILIZERS SHOULD BE USED WITH WATER WHICH HAS MAX TEMPERATURE 25 DEGREE CELSIUS. BUT ACTUAL WATER TEMPERATURE IS GREATER 65 DEGREE CELSIUS. BOTH MODELS OF TOTTNAUER STERILIZERS HAVE SAME PUMPS WHICH DON'T MEET STERILIZERS DESIGN REQUIREMENT. PUMP MODEL IS E, TYPE (B)(4).