FDA Adverse Event
Malfunction
Summary report: N
TABLE TOP STERILIZER
MDR report key: 7818699
·
Received August 24, 2018
Report
- Report Number
- MW5079403
- Event Type
- Malfunction
- Date Received
- August 24, 2018
- Date of Event
- August 8, 2018
- Report Date
- August 14, 2018
- Manufacturer
- TUTTNAUER USA CO., LTD
- Product Code
- FLE
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Description of Event or Problem · 1
DURING REGULAR CYCLE, THE TABLE TOP STERILIZERS FAILED SAFE. EVALUATION OF THEM SHOWED THAT WATER PUMP IS DEFECTIVE. THE PUMP IS FURNISHED AS PART OF STERILIZERS SHOULD BE USED WITH WATER WHICH HAS MAX TEMPERATURE 25 DEGREE CELSIUS. BUT ACTUAL WATER TEMPERATURE IS GREATER 65 DEGREE CELSIUS. BOTH MODELS OF TOTTNAUER STERILIZERS HAVE SAME PUMPS WHICH DON'T MEET STERILIZERS DESIGN REQUIREMENT. PUMP MODEL IS E, TYPE (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 656464 | TABLE TOP STERILIZER | STERILIZER, STEAM | FLE | TUTTNAUER USA CO., LTD | EZ 11 PLUS | ||
| 656465 | TABLE TOP STERILIZER | STERILIZER, STEAM | FLE | TUTTNAUER USA CO., LTD | 3870EA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |